ligo-ex ligo-ds
  Richardson Lab Experimental Log, Page 2 of 11  Not logged in ELOG logo
ID Date Author Type Category Subject Text Attachments
  494   Sun Dec 22 20:43:35 2024 LuisUpdateVACMR Testing[Luis, Luke]
        RGA Volume:
T = 23°C, P = 3.0e-9 Torr
        Main Volume:
 2024-12-13_PostBake2_ArO_Degas_overlay.png2024-12-22_MR-Bake1_ArO.png 
  493   Thu Dec 19 12:00:44 2024 LukeUpdateVACEnd of bakeThe
vacuum chamber has stopped baking. 
Current state as of 11:40, 12/19/2024:
  
  492   Wed Dec 18 15:41:56 2024 TylerUpdateTCSFROSTI RIN[Tyler]


I have begun the (hopefully)
final RIN measurements at 3:30 PM today.
If you need to go into the cleanroom at any
  
  491   Tue Dec 17 13:53:53 2024 LuisUpdateVACVacuum system status [Luis]

 The vacuum system is being
baked. I disconnected the digital gages and
the magnets, opened the Ar leak and started
  
  490   Mon Dec 16 21:32:21 2024 LuisUpdateVACMR Testing[Luis, Luke]

The MR material was
placed in the vacuum chamber following standard
procedures to ensure the vacuum remained
 IMG_6011.HEIC 
  489   Sat Dec 14 07:58:25 2024 LuisUpdateVACVacuum sytem staus and Viton O-rings  [Luis]
        RGA Volume:
T = 23°C, P = 1.71e-9 Torr
        Main Volume:
 2024-12-12_PostBake2_ArO_overlay.png 
  488   Thu Dec 12 13:06:39 2024 LuisUpdateVACRGA Scan and vacuum system updates    [Luis, Luke]
    
        RGA Volume:
T = 21°C, P = 1.74e-9 Torr
 2024-12-12_PostBake2_ArC.png2024-12-12_PostBake2_ArO_overlay.png 
  487   Wed Dec 11 16:06:05 2024 LuisUpdateVACRGA Status[Luis]

 I rebooted Spica and I
was able to connect the RGA again. After
Connecting the RGA, I left the filament on.
  
  486   Wed Dec 11 13:39:30 2024 TylerUpdateElectronicsRIN Update 12/11/2024[Tyler]


Attached below are updated
ASD plots for the FROSTI RIN measurements.
The parameters set for this are the following:
 Updated_ASD_plots.pdf 
  485   Thu Dec 5 10:31:13 2024 Luis UpdateVACRGA Scan and RGA updates



   
    


 2024-12-04_PostBake2_ArO.png 
  484   Wed Dec 4 16:01:23 2024 ShaneUpdateCDSVac serial interfacing troubleshooting updateUpdate on the issue with serial interfacing
for the TV 551 turbo pump: Tried a new test
today, to see if the commands being sent
  
  483   Mon Dec 2 13:35:45 2024 Luke UpdateGeneralClean and bake of Viton O-rings[Luke, Luis]
We couldn't access the bake log on
the wiki. (see attached) So we looked in
 image_(2).png 
  482   Mon Dec 2 00:11:23 2024 MichaelUpdateInterferometer SimulationsStatus Update on Toroidal Mirrors Project WorkAs it has been significant time from my last
update on this project, I compiled a few
notes about what I have been working on,
  
  481   Sat Nov 30 12:41:42 2024 LukeUpdateVACVacuum bake endThe
vacuum chamber has stopped bakeing. 
Current state as of 12:10, 11/30/2024:
  
  480   Wed Nov 27 16:23:58 2024 Luke UpdateVACStarted bake[Luke, Luis]
The
vacuum chamber is currently baking. 
  
  479   Wed Nov 27 13:48:01 2024 TylerUpdateElectronicsRIN Update[Tyler]

 
I downgraded the Red Pitaya
back to OS 2.00-18 due to runtime errors
during measurement. Once I did this, the
 New_ASD_plots.pdf 
  478   Wed Nov 27 12:27:45 2024 Luke SummaryVACSummary of vacuum work[Luke, Luis]
Update of current state of vacuum
chamber
 2024-11-27_PostBake1_ArO.png 
  477   Wed Nov 20 12:59:37 2024 TylerUpdateElectronicsRIN UpdateI went ahead and compared the noise spectrum
of the ADC to those of the photodetectors
with the FROSTI on and off. As discussed
 ASD_plots.pdf 
  476   Wed Nov 20 12:31:24 2024 LuisUpdateVACLeak test RGA Scan after vacuum repairs[Luke, Luis]
After the gasket replacement performed
on Monday (11/18/24) we let the vacuum pump
 2024-11-20_PostBake1_ArO.png 
  474   Tue Nov 19 17:10:34 2024 Luke UpdateVACLeak checkChamber temperature: 23C
Chamber Initial Pressure: ~1.8e-8
main, ~3.5e-8 rga
 image.pngIMG_5116.jpegIMG_5117.jpeg 
  473   Thu Nov 14 11:58:05 2024 ShaneUpdateCDSTurbo pump connection troubleshootingAttempted connection to the TV551 turbo pump
through Varian/Agilent's software on spica.
Successful connection established, and we
  
  472   Thu Nov 14 10:46:20 2024 Luke SummaryCleanroomCleanroom cleaning 
[Luke, Luis, Michael]
cleaning cleanroom and particle count
 partical_count_final.png 
  471   Wed Nov 13 14:06:00 2024 Luke UpdateGeneral slides   
  470   Wed Nov 13 14:03:32 2024 TylerUpdateElectronicsRIN Update[Tyler]


We've added two low-pass filters
in hopes of reducing any potential aliasing
that may be introducing additional noise
 IMG_0463.jpgCH0_CH1_ASDs.pdf 
  469   Wed Nov 6 13:48:55 2024 LiuUpdateInterferometer SimulationsQN modeling updateSlides   
  468   Tue Nov 5 11:08:40 2024 Luis MartinUpdateVACISO Test
    [Luis, Luke, Ma Michael]
    
        The isolation test
was conducted on the vacuum system. Every
  
  467   Tue Nov 5 09:56:04 2024 Luis MartinUpdateClean & BakeMR Bake    [Luke, Luis]
    
        On Friday, 11/1/24,
the MR was unpackaged and cleaned under the
  
  466   Tue Oct 29 16:37:35 2024 TylerUpdateElectronicsRed Pitaya OS Update The Red Pitaya ecosystem has been upgraded
to OS 2.00-35, with a key feature
being greater freedom in adjusting the sampling
  
  465   Fri Oct 25 10:56:30 2024 ShaneUpdateCDSTiming Chassis fixed and reinstalled


[Ma, Shane]

We
replaced the 1A breaker in the timing chassis
today with a 4A one, and tested that all
 successful_timing.jpeg 
  Draft   Wed Oct 23 12:33:54 2024 Luke Update Zernike calculation updatePowerPoint
slides new

PowerPoint
  
  463   Wed Oct 23 12:08:36 2024 Luis MartinUpdateVACRGA Scan[Luke, Luis, Mary, Ma]
 On 10/22/24 Luke, Ma, Mary and myself
ran a RGA scan, the results are displayed
 2024-10-22_PostBake1_ArO.png 
  462   Wed Oct 23 11:48:06 2024 Luis MartinUpdateVACLeak Test Results[Luke, Luis, Mary, Ma]
 On 10/22/24 Luke, Ma, Mary and myself
ran a leak test, the results are displayed
  
  461   Thu Oct 17 13:24:23 2024 ShaneUpdateCDSTiming Chassis issue identified



[Ma,
Luis, Shane]

Working theory for the
timing chassis issues had been that the 1A
 IMG_6462.jpegIMG_6464.jpeg 
  460   Wed Oct 16 14:13:31 2024 LiuUpdate FROSTI with non uniform absorption scattering sourcesSlides   
  459   Wed Oct 16 13:58:59 2024 Luis MartinUpdateVACRGA ScanThe following figure displays data acquired
on 10/09, and it shows that we are no longer
below the cleanliness standard. Also, the
 2024-10-09_PostBake1_ArO.png 
  458   Tue Oct 15 15:27:38 2024 Xuesi MaUpdateELOGTiming Chassis Update
I conducted separate tests on the '5015'
and '3010a'. When powered individually, the
'5015' outputs a signal at 33.55 MHz with
 12x 
  457   Wed Oct 9 13:52:31 2024 TylerUpdateTCSRIN Update 10/09/2024 I tried adjusting the gain settings on
the photodetectors to check if this would
help improve the RIN spectra measurements.
 RIN_plots_CH0.pdf 
  456   Thu Oct 3 19:35:04 2024 ShaneUpdateCleanroomCleanroom cleaning and particle count


[Luke, Michael,
Cynthia, Mary]


cleaning cleanroom
and particle count
  
  455   Wed Oct 2 14:49:59 2024 LiuUpdate FROSTI ETM actuationProposed FROSTI ETM actuation on the HOM7
resonance.  

Animation
  
  454   Wed Oct 2 13:38:00 2024 CynthiaUpdateInterferometer Simulationscavity scan with averaged O4 ETM coating as plume   O5plumevsCoating.pngO5CoatAvgpt5.pngO5AVgcoatingFROSTIpt5.PNG 
  453   Wed Oct 2 12:20:43 2024 TylerUpdateTCSUpdated FROSTI Analysis 10/02/2024Attached below are updates on current FROSTI
prototype analysis.
 FROSTI_Analysis_Updates.pdf 
  452   Wed Oct 2 12:05:42 2024 Luke Update Ringheater modeling UpdatePower
point slides
  
  451   Wed Oct 2 10:31:46 2024 Xuesi MaConfiguration Group Meeting Slides 10/2/2024
Group Meeting slides for Non-deterministic
Heater Response.
 Group_Meeting_10_2.pdf 
  Draft   Tue Oct 1 19:41:23 2024 ShaneUpdateCDSCyMAC Timing Chassis Issue[Shane, Luis]


Summary of the issue we've
been having with the timing chassis: when
we connect only the Valon 5015 to power,
 IMG_5917.jpegIMG_5803.jpegIMG_5804.jpeg 
  448   Wed Sep 25 09:06:09 2024 Luis MartinUpdateVACRGA Scans



   
    
 24overlay.png24overlayNormalized.png 
  447   Mon Sep 23 15:11:21 2024 TylerUpdateScripts/ProgramsEffective Emissivity Analysis and RIN Update A quick update on the effective emissivity
analysis for the CIT FROSTI testing: 

 
I was able to (roughly) match
 OPD_Plot.pngTemp_emiss_plot.pngrin_photodectors.pngRIN_setup.jpg 
  446   Mon Sep 23 15:00:41 2024 Cynthia UpdateInterferometer Simulationscorrections on assumption for O5 coatingThe document recording ETM07 and ETM08 coating
(T2300093) was discussed: 

the current
plume files are not necessarily correct in
 Uncoated_S1_ETM07-v2.pdf 
  444   Mon Sep 23 08:29:49 2024 LukeUpdateInterferometer SimulationsRingheater update
Ringheater
Update
If the link does not work here is
 20240923_Update.pptx 
  443   Mon Sep 16 15:18:59 2024 Cynthia UpdateInterferometer Simulationsa breakdown of FROSTI and thermal effects and the resulting cavity scan    ligo_update_9_16_(1).pdf 
  442   Mon Sep 16 14:59:51 2024 TylerUpdateTCSFLIR RIN Update 09/16
Below is the dark noise spectrum of the
Red Pitaya, which was measured over the course
of a weekend. Additionally, I have successfully
 adc_noise_floor_photodectors.png 
ELOG V3.1.3-7933898