ligo-ex ligo-ds
  Richardson Lab Experimental Log, Page 1 of 12  Not logged in ELOG logo
ID Date Authordown Type Category Subject
  12   Mon Nov 28 18:10:23 2022 shaneUpdateELOGParticle counts in the clean room
Particle count stats for the clean room Nov 28, 2022: Took 10 sample runs in each of 5 regions in the clean room (5 runs per region with a person inside the clean room for the measurement, and 5 runs per region without anyone in the clean room for the measurement), for a total of 50 samples taken. Sample time was 60 seconds. Overall clean room average particle count for the size ranges are as follows: 0.3 micrometers- 3405.76 (room occupied), 974.92 (room empty) 0.5 micrometers- 409.72 (room occupied), 409.72 (room empty) 1.0 micrometers- 1102.2 (room occupied), 282.6 (room empty) 2.5 micrometers- 692.32 (room occupied), 183.68 (room empty) 4.0 micrometers- 254.28 (room occupied), 84.72 (room empty) 5.0 micrometers- 141.24 (room occupied), 84.72 (room empty) 7.0 micrometers- 56.48 (room occupied), 84.72 (room empty) 10.0 micrometers- 42.36 (room occupied, 42.36 (room empty) More statistics (including individual stats on the 5 regions within the clean room) attached.
Attachment 1: clean_room_particle_counts_11_28_-_Sheet1.pdf
clean_room_particle_counts_11_28_-_Sheet1.pdf
  19   Fri Feb 3 13:04:04 2023 shaneSummaryGeneralclean room particle counts 1/25/23
Clean room count graphs for each zone (as of January 25, 2023) attached
Attachment 1: cleanroomcountsJan25.pdf
cleanroomcountsJan25.pdf
  71   Wed Apr 12 16:17:50 2023 shaneUpdateGeneralParticle Counter moved
With the added height of the new docking station, the particle counter no longer fits under the Windows monitor in the electronics rack and has been moved to the desk in the corner of the lab.
Attachment 1: IMG_6693.HEIC
  178   Fri Jul 28 13:39:39 2023 shaneUpdateCleanroomfull 5 zone cleanroom particle count
Here's the 5 zone cleanroom measurement from today.
Attachment 1: download.png
download.png
  193   Mon Aug 7 13:58:46 2023 shaneUpdateCDSinternal power supply cables done for AI chassis
Finished assembling and installing the aLigo chassis internal power supply cables in the anti-imaging chassis today. Tested for continuity and everything looked good. Chassis lid still secured with two temporary screws because I'm not sure what the correct size is.
Attachment 1: IMG_8674.jpeg
IMG_8674.jpeg
  260   Thu Nov 9 19:06:46 2023 shaneUpdateCleanroom5 zone cleanroom measurement
Here's today's five zone measurement, taken ~5 hours after the cleaning. Zone 5 (closest to fire cabinet and back wall) is a bit over the limit in all size ranges (about 1500 particles in 1.0u size range over), and zone 2 is slightly over the limit in the 1.0u range, but everything else is under the requirement.
Attachment 1: 23.jpeg
23.jpeg
  262   Fri Nov 10 14:16:04 2023 shaneUpdateCleanroomnew cleanroom zone diagram
since the vac system has been moved and made permanent, here's a new diagram of the cleanroom zones. The zones themselves haven't moved, but the diagram has been updated to reflect the new set up of the cleanroom.
Attachment 1: cleanroom_zones_nov2023.jpg
cleanroom_zones_nov2023.jpg
  309   Thu Jan 18 18:06:49 2024 shaneUpdateCleanroomcleanroom 5 zone particle count measurement
Here's today's full five zone measurement of the cleanroom
Attachment 1: 24.png
24.png
  315   Fri Feb 2 19:18:24 2024 shaneUpdateCleanroomcleanroom cleaning and particle count
[Aiden, Shane, Luis, Luke]

cleaning cleanroom and particle count

  • 5:10 pm: ran zero count test on particle counter
  • 5:14 pm: started particle count

    NOTE: initial counts are significantly above allowed limit. ~10 times larger than ISO class 5 standard in the smallest size range. Realized partway through ceiling tile changes that the power/lights for some of the hepa fans in the ceiling frame had been turned off. May explain this high number. Have since turned fans back back on.

    • zone 3:
      • 0.3 u: 109208
      • 0.5 u: 16337
      • 1.0 u: 581
    • zone 4:
      • 0.3 u: 51340
      • 0.5 u: 9228
      • 1.0 u: 581
  • 5:33 pm: started replacing ceiling tiles
  • 6:25 pm: began surface check and wipedown, including softwalls
  • 6:33 pm: started vacuuming the floor
  • 6:45 pm: finished vacuuming the floor
  • 6:45 pm: started mopping the floor
  • 6:50 pm: finished mopping the floor
  • 6:52 pm: started cleaning the buckets
  • 6:53 pm: started mopping with IPA wipes
  • 6:58 pm: finished mopping with IPA wipes
  • 6:59 pm: changed sticky floor mats
  • 7:00 pm: started particle count
    • zone 3:
      • 0.3 u: 5404
      • 0.5 u: 1413
      • 1.0 u: 41
    • zone 4:
      • 0.3 u: 1704
      • 0.5 u: 332
      • 1.0 u: 166
  318   Wed Feb 7 12:43:22 2024 shaneUpdateCleanroomcleanroom particle counts (full 5 zone measurement)
Here's today's full 5 zone measurement of the cleanroom. Everything is back under the requirement.
Attachment 1: 24.png
24.png
  365   Tue May 7 20:34:31 2024 shaneUpdateCDSfrosti MEDM screen update
[Luis, Shane] Here is the updated MEDM screen, with new orientation and updated labeling to reflect the actual positions of the heater elements. Note that indices start at T0 to be consistent with simulink model, though in the previous elog for the FROSTI layout the heater elements are labeled 1-8. Also, we finally learned how to take a screenshot on debian.
Attachment 1: FROSTIMEDMMay7.2024.png
FROSTIMEDMMay7.2024.png
  505   Tue Jan 28 13:30:46 2025 shaneUpdateCDSSerial comms testing update
Attempted to communicate with the smaller vac pump (TwissTor 74) via Agilent software today. Was able to communicate with the pump via python serial connection code as usual, with no issues. When using the manufacturer software though, it was unable to 'locate' the pump and failed to connect to it. So manufacturer software works for the big (older) pump, but not the smaller, newer one. Unclear if this is somehow related to the fact that using the manufacturer software for the bigger pump involved manually connecting its controller to spica with a db9, and there was no obvious way to do that for the smaller pump's controller.
  546   Thu Apr 3 13:23:37 2025 shaneUpdateCDSUPS interfacing
Update on UPS serial interfacing: the driver for the UPS webcard has been installed on spica (for UPS1 in room 1119) and logrus (for UPS2 in room 1129). Static IP addresses have been assigned for both and noted on the lab wiki. Neither is yet able to connect, as the initial ip address webpage connection test failed for both. In the process of troubleshooting now.
  451   Wed Oct 2 10:31:46 2024 Xuesi MaConfiguration Group Meeting Slides 10/2/2024

Group Meeting slides for Non-deterministic Heater Response.

Attachment 1: Group_Meeting_10_2.pdf
Group_Meeting_10_2.pdf
  458   Tue Oct 15 15:27:38 2024 Xuesi MaUpdateELOGTiming Chassis Update

I conducted separate tests on the '5015' and '3010a'. When powered individually, the '5015' outputs a signal at 33.55 MHz with an amplitude of 608 mV. It draws 1 A of current from the power source. The input signal for the '3010a' is 33.54 MHz with an amplitude of 670 mV (peak-to-peak) and a 15 mV DC offset. The output signal from channel 1 is a 65.5 kHz square wave with an amplitude of 3.28 V. The '3010a' draws 0.1 A of current.

Both the '5015' and '3010a' work fine when powered separately. However, when both are powered together, the power source behaves as if there is a short circuit. The current theory is that the switch or breaker is tripping, as it has a 1 A current rating. Since the combined current demand of both devices exceeds 1 A, this may be causing the issue.

Slides for 10/16/2024 Group Meeting

Attachment 1: 20241015_134152.mp4
Attachment 2: 20241015_133118.png
20241015_133118.png
Attachment 3: 20241015_133124.png
20241015_133124.png
Attachment 4: 20241015_133136.png
20241015_133136.png
Attachment 5: 20241015_133207.png
20241015_133207.png
Attachment 6: 20241015_133543.png
20241015_133543.png
Attachment 7: 20241015_133550.png
20241015_133550.png
Attachment 8: 20241015_133555.png
20241015_133555.png
Attachment 9: 20241015_133300.png
20241015_133300.png
Attachment 10: 20241015_133643.png
20241015_133643.png
Attachment 11: 20241015_134202.png
20241015_134202.png
Attachment 12: 20241015_135353.png
20241015_135353.png
  503   Mon Jan 27 11:29:38 2025 Xuesi MaUpdate Heater Element Test

[Ma, Cece, Luke, Mary, Shane]

On Friday (Jan 24), we installed the heater elements on the stand. The heater elements are arranged from 1 to 8, oriented from right to left as shown in Attachment 1. Each wire has been labeled according to its corresponding element number and type (e.g., RTD connections, heater connections).

Note: We currently do not have enough PEEK zip ties, so standard zip ties have been used temporarily. These must be replaced with PEEK zip ties before the setup is placed in the vacuum chamber.

Attachment 1: 20250124_165434.png
20250124_165434.png
Attachment 2: 20250124_165342.png
20250124_165342.png
Attachment 3: 20250214_132811.jpg
20250214_132811.jpg
  504   Mon Jan 27 23:35:28 2025 Xuesi MaUpdate  

[Ma]

Installed all the pins to the peek DB 25 connectors

Attachment 1: 20250127_142647.jpg
20250127_142647.jpg
Attachment 2: 20250127_153454.jpg
20250127_153454.jpg
Attachment 3: 20250127_154516.jpg
20250127_154516.jpg
  506   Fri Jan 31 15:03:09 2025 Xuesi MaUpdate Heater Element Circuit Check

[Ma] Wed 1/29/2025

No short circuit between heater element ✓

No short circuit to ground on any pin ✓

No short circuit between connectors ✓

Heater Number Power Resistor before (Ohm) Power Resistor now (Ohm) RTD Resistor before (Ohm) RTD Resistor now (Ohm)
1 73.6 73.1 81.8 81.3
2 70.4 69.6 82.1 81.6
3 71 70.5 84.5 84
4 71.5 71 80 79.4
5 70.5 70.2 81.7 81.2
6 72 71.6 79.4 78.7
7 69.2 69 78.2 77.5
8 71.1 70.6 84.2 83.6
  508   Mon Feb 3 13:17:21 2025 Xuesi MaUpdate  

[Ma]

changed all zip ties to peek zip ties, and grouped wires together. The setup is ready to go into the chamber.

Attachment 1: 20250203_123153.jpg
20250203_123153.jpg
Attachment 2: 20250203_123213.jpg
20250203_123213.jpg
  514   Tue Feb 18 10:01:07 2025 Xuesi MaUpdate  

[Ma, Pooyan, Tyler]

On Friday, we connected the vacuum chamber with the Cymac.

Attachment 1: IMG_7916.jpeg
IMG_7916.jpeg
Attachment 2: IMG_7915.jpeg
IMG_7915.jpeg
Attachment 3: IMG_7917.jpeg
IMG_7917.jpeg
  515   Tue Feb 18 10:02:16 2025 Xuesi MaUpdate Group Meeting Slides 2/18/2025

https://docs.google.com/presentation/d/1WiV2VqS0BzXNCK6VYYQ-Ty8xlHnzXatQaFbMf1-0rsY/edit?usp=sharing

  525   Tue Feb 25 10:02:32 2025 Xuesi MaUpdate  

1424039912.625576 2025/02/19 22:38:14 UTC Time start

24V 2.8A right after start all 8 elements

1424044804.902443 2025/02/19 23:59:46 UTC Time stop

24V 1.8A right before stop all 8 elements

0.1A right before start and right after stop


note: turned on briefly to check current right before stop


2/20 RGA Scan


spikes??


1424129207.857777 2025/02/20 23:26:29 UTC Time start

24V 2.9A right after start all 8 elements

1424140789.856096 2025/02/21 02:39:31 UTC Time stop

24V 1.7A right before stop all 8 elements

0.1A right before start and right after stop


2025/02/21 02:43:19 UTC

Main chamber pressure: 1.54e-8

RGA chamber pressure:5.06e-9


spikes are due to loose connection between connectors.


1424210410.173863 2025/02/21 21:59:52 UTC Time start

24V 2.9A right after start all 8 elements

1424218357.96404 2025/02/22 00:12:19 UTC Time stop

24V 1.7A right before stop all 8 elements

each elements: 0.3A (0.2A increment)(all)

0.1A right before start and right before stop


2/24 RGA Scan


1424467918.129082 2025/02/24 21:31:40 UTC Time start

12V 2A right after start all 8 elements

1424478936.635821 2025/02/25 00:35:18 UTC Time stop

12V 1.6A right before stop all 8 elements

each elements: 0.4A (0.2A increment)(all)

0.2A right before start



rise time: A(1-exp(-t/tau))+B

fall time: Aexp(-t/tau) +B

Attachment 1: resistance_vs_time_all_channels_all_time.pdf
resistance_vs_time_all_channels_all_time.pdf
Attachment 2: temperature_vs_time_all_channels_all_time.pdf
temperature_vs_time_all_channels_all_time.pdf
Attachment 3: resistance_vs_time_rise_time.pdf
resistance_vs_time_rise_time.pdf
Attachment 4: resistance_vs_time_fall_time.pdf
resistance_vs_time_fall_time.pdf
  528   Tue Mar 4 03:05:42 2025 Xuesi MaUpdateELOGLOG

1424564097.243212 2025/02/26 00:14:39 UTC Time start

12V 2A right after start all 8 elements

1424564478.079428 2025/02/26 00:21:00 UTC Time stop

12V 1.8A right before stop all 8 elements

1424564832.211584 2025/02/26 00:26:54 UTC Time start

12V 1.9A right after start all 8 elements

1424565208.359935 2025/02/26 00:33:10 UTC Time stop

12V 1.8A right before stop all 8 elements

1424565573.565066 2025/02/26 00:39:15 UTC Time start

12V 1.8A right after start all 8 elements

1424565931.242394 2025/02/26 00:45:13 UTC Time stop

12V 1.7A right before stop all 8 elements

1424566292.67104 2025/02/26 00:51:14 UTC Time start

12V 1.8A right after start all 8 elements

1424566648.619952 2025/02/26 00:57:10 UTC Time stop

12V 1.7A right before stop all 8 elements

1424566996.312246 2025/02/26 01:02:58 UTC Time start

12V 1.8A right after start all 8 elements

1424567381.748943 2025/02/26 01:09:23 UTC Time stop

12V 1.7A right before stop all 8 elements

1424567756.528736 2025/02/26 01:15:38 UTC Time start

12V 1.7A right after start all 8 elements

0.2A right before start


spikes!?


1424643001.864687 2025/02/26 22:09:43 UTC

change c_0(VEXC0 & VCXC0) to 2V (why is it on 5V ?)


2025/02/26 22:18:51 UTC

Main chamber pressure:5.92e-9

RGA chamber pressure:1.98e-9


1424650528.240947 2025/02/27 00:15:10 UTC Time start (increase voltage)

24V 2.1A right after start all 8 elements


disconnect and reconnect (exc 1-4)(out 9-12) & (exc 5-8)(out 13-16)


2025/02/27 19:34:26 UTC

Main chamber pressure:1.34e-8

RGA chamber pressure:4.33e-9

Main chamber temp: 29

RGA chamber temp:29


2025/02/28 18:02:29 UTC

Main chamber pressure:1.05e-8

RGA chamber pressure:3.53e-9

Main chamber temp: 29

RGA chamber temp:30


1424801097.374902 2025/02/28 18:04:39 UTC Time stop

24V 1.8A right before stop all 8 elements

0.1A right after stop


1424827460.71372 2025/03/01 01:24:02 UTC Time start

24V 2.9A right after start all 8 elements

0.1A right before start


2025/03/01 01:25:34 UTC

Main chamber pressure:5.98e-9

RGA chamber pressure:2.06e-9

Main chamber temp:27

RGA chamber temp:27


2025/03/03 20:28:03 UTC

Main chamber pressure:7.8e-9

RGA chamber pressure:2.64e-9

Main chamber temp:27

RGA chamber temp:27

  529   Tue Mar 4 03:33:18 2025 Xuesi MaUpdate Update

Spikes Appear Again, need to address it systematically.

Power on and off before reaching steady state ✔

At 12V, the rise and fall time of heater elements are different from 24V.

Initial guess is due to temperature in the chamber. However, it does not seem to be the case, 24V with 30c have the same hall time as 24V at lower temperature.

Attached are some graphs for rise and fall time

After a weekend of powering on, the main chamber pressure stabilized in the UHV region.

Temperature in the chamber seems also not to change.

Attachment 1: resistance_vs_time_all_channels_all_time.pdf
resistance_vs_time_all_channels_all_time.pdf
Attachment 2: resistance_vs_time_rise_time.pdf
resistance_vs_time_rise_time.pdf
Attachment 3: resistance_vs_time_fall_time.pdf
resistance_vs_time_fall_time.pdf
Attachment 4: resistance_vs_time_rise_time_12V.pdf
resistance_vs_time_rise_time_12V.pdf
Attachment 5: resistance_vs_time_fall_time_12V.pdf
resistance_vs_time_fall_time_12V.pdf
Attachment 6: resistance_vs_time_fall_time_24V_30c.pdf
resistance_vs_time_fall_time_24V_30c.pdf
  539   Tue Mar 11 10:43:57 2025 Xuesi MaUpdate  

Continue investigation in the spikes

Pulsed ADC with a function generator and find no spikes. Rules out ADC for causing the spikes

Loopback test that bypasses the FROSTI Chassis shows spikes (spikes happened on all channels at the same time)

Next step is to bypass the AI Chassis to find the source of the spikes

Attachment 1: 20250310_171936.jpg
20250310_171936.jpg
  544   Mon Mar 31 15:41:56 2025 Xuesi MaUpdate  

The result from AI Chassis bypass test showed that the AI Chassis may be the problem. There are no spikes from DAC's direct output.

All chassis except timing chassis are turned off. Power supplies for 24V, 18V, and -18V are turned off

The AI chassis has been taken out of the rack for further inspection

Refer to:

https://dcc.ligo.org/LIGO-E2300117

https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?.submit=Identifier&docid=D2300124&version=

  547   Mon Apr 7 13:28:49 2025 Xuesi MaUpdate  

Test 1: Adapter Board Connected to Filter Board

  • Setup: AI Chassis is powered on. Signal is measured directly from the adapter board. All other ports on the adapter board are connected to the filter board.
  • Observation: Two distinct spikes observed at the beginning of the measurement.

Test 2: Adapter Board Disconnected from Filter Board

  • Setup: AI Chassis remains powered on. Signal is again measured directly from the adapter board. This time, the remaining ports on the adapter board are not connected to the filter board.
  • Observation: Multiple spikes observed, distributed evenly across the entire measurement.

Test 3: Isolation Test with Spare Adapter Board (ongoing)

  • Setup: Suspecting the original adapter board may be faulty, a spare adapter board is used for comparison. Signal is measured directly from this spare board.

Update, 04/08/25, Tue 17:30

Took the old adapter board out of the AI chassis and and used it to connect DAC to AA chassis. If no spikes are seen, it means that the glitches are not originating from this board.

Also checked al the previous spikes using the raw data (65536 Hz sample rate). The duration of the glitches are ~1 second, despite the previous guess that they are happening in the Milli-second scale.

Attachment 1: 20250403_151946.jpg
20250403_151946.jpg
Attachment 2: ELOG_Spike_adapter_filter_on.png
ELOG_Spike_adapter_filter_on.png
Attachment 3: 20250404_143624.jpg
20250404_143624.jpg
Attachment 4: ELOG_Spike_adapter_isolate.png
ELOG_Spike_adapter_isolate.png
Attachment 5: 20250407_105846.jpg
20250407_105846.jpg
  553   Tue Apr 15 11:13:43 2025 Xuesi MaUpdate  

Test 4: AI Chassis ground isolation

  • Setup: AI Chassis isolated from the ground with readout directly connected to adapter board
  • Observation: No spike observed.

Test 5: Ribbon cable check

  • Setup: AI Chassis remains isolated from the ground. The signal is measured after the ribbon cable. (Attachment 3)
  • Observation: No spike observed.

Test 6: Single ground connection check

  • Setup: AI chassis powered on but remains isolated from the frame. The signal is measured from the output of the AI Chassis. (Attachment 1 ,2)
  • Observation: No spike observed.
Attachment 1: 20250414_120428.jpg
20250414_120428.jpg
Attachment 2: 20250414_120423.jpg
20250414_120423.jpg
Attachment 3: 20250411_163105.jpg
20250411_163105.jpg
  559   Tue Apr 22 12:08:46 2025 Xuesi MaUpdate  

Test 7: Reconnect ribbon cables

  • Setup: Reconnect all the ribbon cables, and turned on the AI Chassis while isolated rack
  • Observation: No spike observed.

Test 8: Reconnect to rack

  • Setup: Reconnect the AI Chassis to the rack.
  • Observation: No spike observed.

Test 9: Restore to original position.

  • Setup: CLose up the AI Chassis, and put it back to its original location.
  • Observation: No spike observed.
Attachment 1: 20250415_161755.jpg
20250415_161755.jpg
Attachment 2: 20250416_174640.jpg
20250416_174640.jpg
Attachment 3: 20250422_105614.jpg
20250422_105614.jpg
  567   Wed Apr 30 17:33:14 2025 Xuesi MaUpdateVACStarted bake

[Luke, Ma, Tyler]

The vacuum chamber is currently baking. 

Current state as of 5:00

The gate valve is open, and the filament of the RGA is on. Argon leak is opened

The temperatures are as follows:

PID right barrel upper: 99°C,  RGA volume: 120°C

PID left barrel lower: 123°C, Lid: 92°C

  568   Fri May 2 12:34:02 2025 Xuesi MaUpdateVACStopped bake

The vacuum chamber has stopped baking. 

Current state as of 12:30, 5/2/2025:

The gate valve is open, and the filament of the RGA is on.

The temperatures were steady at:

PID right: barrel upper: 108°C,  RGA volume: 120°C

PID left: barrel lower: 120°C, Lid: 91°C

  572   Tue May 6 13:19:31 2025 Xuesi MaUpdate  
Attachment 1: resistance_vs_time_RV_rise_time_24.pdf
resistance_vs_time_RV_rise_time_24.pdf
Attachment 2: resistance_vs_time_all_channels_all_time_RV.pdf
resistance_vs_time_all_channels_all_time_RV.pdf
  576   Tue May 13 12:35:01 2025 Xuesi MaUpdate  
Attachment 1: resistance_vs_voltage_all.pdf
resistance_vs_voltage_all.pdf
Attachment 2: resistance_vs_voltage_overlay.pdf
resistance_vs_voltage_overlay.pdf
Attachment 3: resistance_vs_time_all_channels_all_time_RV.pdf
resistance_vs_time_all_channels_all_time_RV.pdf
  578   Thu May 15 12:03:41 2025 Xuesi MaInfrastructureElectronicsNotice

All Sorensen power supplies are turned off. The Cymac system is down (All Chassis are down)

If you need more information or if you need to turn them back on please contact Xuesi Ma.

  579   Mon May 19 10:59:14 2025 Xuesi MaInfrastructure  

Status Update

Previous Status:

All Sorensen power supplies were turned off. The Cymac system was down (all chassis were offline).

If you need more information or assistance in turning them back on, please contact Xuesi Ma.


Update: The systems are now back online.

Note: The Sorensen power supplies’ voltage can now be controlled via channel VEXC8 of the DAC.

If scripting is required, the ezca Python package is highly recommended for interfacing and control.

  580   Wed May 21 14:40:38 2025 Xuesi MaInfrastructureCleanroomCleanroom Temperature

Date and Time: Around 2:20 PM on May 21, 2025

Location and Temperature:

  • Back of the room, around the working station: 85.5 °F
  • Front of the room, around the doorway: 82.3 °F
Attachment 1: 20250521_142051.jpg
20250521_142051.jpg
Attachment 2: 20250521_142220.jpg
20250521_142220.jpg
  583   Tue May 27 12:52:22 2025 Xuesi MaUpdate  

Using the configuration in section 2.11.1 in the Sorensen installation and operation manual (attached below), I'm able to remote control the Sorensen power supply.

I use DAC VEXC8 as the external voltage source and a breakout board to match the configuration of connector J3.

I have recorded 7 cycles (attached below), and am now working on analyzing the data.

Attachment 1: operating-manual-Sorensen-DCS-programmable-switching-dc-power-supplies-1kw-dcs-e.pdf
operating-manual-Sorensen-DCS-programmable-switching-dc-power-supplies-1kw-dcs-e.pdf
Attachment 2: 20250515_171905.jpg
20250515_171905.jpg
Attachment 3: 20250515_164637.jpg
20250515_164637.jpg
Attachment 4: 20250515_164634.jpg
20250515_164634.jpg
Attachment 5: 20250515_164842.jpg
20250515_164842.jpg
Attachment 6: resistance_vs_time_all_channels_all_time_RV_ladder.pdf
resistance_vs_time_all_channels_all_time_RV_ladder.pdf
  588   Tue Jun 3 13:08:47 2025 Xuesi MaUpdate  

We currently have the parts for a DB9 connector but need parts for a DB25 connector, where did we get the parts for the DB9 connector? We may be able to use the same company for the DB25 connector.

Also need a cable for the connectors. Find a DB9 to DB9 cable in 1129, not sure if I can disassemble it and use just the cable.

The pressure in the vacuum chamber rises to 2.12e-8 torr, which needs baking.

Attachment 1: 934.jpg
934.jpg
Attachment 2: 933.jpg
933.jpg
  589   Tue Jun 3 13:16:53 2025 Xuesi MaInfrastructure  

Date and Time: Around 3:50 PM on June 2, 2025

Location and Temperature:

  • Back of the room 1119, around the working station: 86 °F
  • Front of the room 1119, around the doorway: 84.1 °F
  • Back of the room 1129, around the working station: 78.6 °F
  • Front of the room 1129, around the doorway: 78.3 °F
  • In the hallway in front of room 1119: 75.2 °F
Attachment 1: 928.jpg
928.jpg
Attachment 2: 929.jpg
929.jpg
Attachment 3: 931.jpg
931.jpg
Attachment 4: 930.jpg
930.jpg
Attachment 5: 932.jpg
932.jpg
  591   Tue Jun 10 12:13:00 2025 Xuesi MaUpdate  

By measuring the resistance of the ladder, I obtained a graph of resistance as a function of voltage.
All the heater elements behave consistently, showing only a small standard deviation.
However, heater element 5 shows a significantly larger standard deviation.
Upon examining the initial ladder graph, I noticed that the resistance of heater element 5 increases following a power outage that resets the system.
At this point, I am unsure why this behavior occurs.

Attachment 1: resistance_vs_voltage.pdf
resistance_vs_voltage.pdf
Attachment 2: resistance_vs_voltage_element_5.pdf
resistance_vs_voltage_element_5.pdf
Attachment 3: resistance_vs_time_all_channels_all_time_RV.pdf
resistance_vs_time_all_channels_all_time_RV.pdf
  594   Tue Jun 17 12:22:43 2025 Xuesi MaUpdate  
CYMAC Remote Control Cable Documentation

CYMAC Remote Control Cable Assembly

Cable Functions

The custom cable assembly serves two primary functions:

1. Jumper Connections (DB25 Side - Connector J3)

The following pins are jumpered on the DB25 connector:

From Pin To Pin
Pin 3 Pin 16
Pin 10 Pin 11
Pin 12 Pin 24
Pin 13 Pin 25
Pin 22 Pin 23

2. Signal Wiring (DB9 to DB25)

The following connections are made between the DB9 and DB25 connectors:

DB9 Pin DB25 Pin Function
Pin 1 Pin 9 Positive side
Pin 6 Pin 12 Negative side
Pin 5 Pin 6 Ground

Note: All connections are verified with a multimeter.

Attachment 1: 20250616_132743.jpg
20250616_132743.jpg
Attachment 2: 20250616_132945.jpg
20250616_132945.jpg
Attachment 3: 20250616_133235.jpg
20250616_133235.jpg
Attachment 4: 20250616_180103.jpg
20250616_180103.jpg
Attachment 5: 20250616_180129.jpg
20250616_180129.jpg
Attachment 6: 20250616_180132.jpg
20250616_180132.jpg
Attachment 7: 20250616_180317.jpg
20250616_180317.jpg
Attachment 8: 20250616_181011.jpg
20250616_181011.jpg
Attachment 9: 20250616_181033.jpg
20250616_181033.jpg
Attachment 10: 20250616_181052.jpg
20250616_181052.jpg
Attachment 11: 20250616_132759.jpg
20250616_132759.jpg
  596   Tue Jun 24 11:50:15 2025 Xuesi MaUpdate  
Heater Element Resistance Test Results

Heater Element Resistance Test Report

Test Overview

Conducted several on-and-off tests on heater elements at 24V, followed by switching heater elements 5 and 8 by changing their connection pins to the ADC.

Resistance Measurements (Ω)

Heater Element Before Switch After Switch Ladder Test Difference (After-Before)
Element 1 153.09577918 153.11447008 153.83373894 +0.0186909
Element 2 149.70967557 149.8175605 150.08915205 +0.10788493
Element 3 157.88314236 157.87136021 157.39696467 -0.01178215
Element 4 146.80153622 146.78826747 147.77923177 -0.01326875
Element 5 151.80888657 152.2204132 150.44919044 +0.41152663
Element 6 148.24375881 148.32846152 146.89161192 +0.08470271
Element 7 144.72286546 144.76879963 145.03627875 +0.04593417
Element 8 151.4564267 149.14240417 149.88675562 -2.31402253

Supporting Graphs

Attached below are the resistance-time graphs for all heater elements:

  • Complete test duration graph
  • Pre-switch period graph
  • Post-switch period graph
  • Resistance voltage graph after removing data from after power outage

Analysis & Conclusion

The switch did not physically change the resistance of the heater elements, suggesting the measurements reflect their true resistances.

Attachment 1: resistance_vs_time_all_channels_channel_switch.jpg
resistance_vs_time_all_channels_channel_switch.jpg
Attachment 2: resistance_vs_time_all_channels_channel_switch_one_cycle_before.jpg
resistance_vs_time_all_channels_channel_switch_one_cycle_before.jpg
Attachment 3: resistance_vs_time_all_channels_channel_switch_one_cycle_after.jpg
resistance_vs_time_all_channels_channel_switch_one_cycle_after.jpg
Attachment 4: resistance_vs_voltage.jpg
resistance_vs_voltage.jpg
  389   Fri Jun 28 10:43:13 2024 XuejunUpdateCleanroomFrosti
[Xuejun, Tyler]

We moved Frosti into the cleanroom and debugged it to make sure everything was working. One of the DB25 pins broke off at the connector for element 1 so it needed to be recrimped. Element 6 short circuited but fixed with adjustment to the heater power pin position. We connected the power and sensor connectors to the power box and recalibrated the RTD sensors.

Attachment 1: frosti.jpg
frosti.jpg
Attachment 2: chassis.jpg
chassis.jpg
  396   Mon Jul 8 10:39:14 2024 XuejunUpdateFLIRAverage Temperature Profile

After taking data for each of the individual heater elements, I imported them into python and overlayed them to produce an average temperature profile. I rotated the 7 of the elements to align with element 1's profile and averaged them out. By setting the range to 28C - 33C (this gave the best visibility of the heating pattern) it gave the profile attached.

Attachment 1: Average-Temperature-Profile.pdf
Average-Temperature-Profile.pdf
  421   Tue Aug 6 13:07:07 2024 XuejunUpdateInterferometer Simulations 
[Xuejun]

The width and location of the measured in-air change in temperature profile has been determined to be 0.045m and 0.137m respectively. Subsequently, a fake irradiance profile was able to be generated that best resembled what the actual irradiance profile could be using this information for testing in COMSOL. The generated irradiance profile that output the most similar change in temperature profile as the measured in-air profile has been included as well as the change in temperature profile it produced on the blackbody screen "test mass" model in COMSOL.

Attachment 1: Thermal_model_image.png
Thermal_model_image.png
Attachment 2: generated_irradiance_profile.png
generated_irradiance_profile.png
  235   Mon Oct 9 11:29:37 2023 TylerUpdateElectronicsPSD/CSD Plot Updates
Continuing from Sophia's SURF Project this summer: Plotting the Power-Spectral Densities (PSDs) and Cross-Spectral Densities of two signals. A continued issue that has been observed in the plots is the random "jump" of one PSD curves. Below, a 1.4 kHz signal is driven with a magnitude of 0.7 Vpp, and connected to the two output ports of the Red Pitaya via an SMC T-Adapter. At the moment, I anticipate that this bug might have something to do with the Real-Time GUI code being used, since this hasn't been observed when running the calculations without it (see below).
Attachment 1: Screenshot_2023-10-09_at_11.20.25_AM.png
Screenshot_2023-10-09_at_11.20.25_AM.png
Attachment 2: welch_csd-2.png
welch_csd-2.png
  243   Mon Oct 23 11:28:05 2023 TylerUpdateCamerasFLIR Camera Code Updates
The past week I've been spending time going through the FLIR code on gitlab. Initially, it had appeared that our measurements with the camera had differed about 2 degrees C from what the thermocouple was giving (Ref ELOG 181). Upon inspection of the FLIR streaming code, I noticed a few issues: 1. The emissivity value wasn't set correctly (was 0.999 originally, should be about 0.95). 2. The ambient temperature was set to about 21.6 C (71 F). 3. The distance given before was about an inch off from what I measured. These three parameters all must be manually specified to calculate the temperature values, and are what I assume caused the larger temperature difference seen before. I've attached a new set of measurements that I took below, where 6 images were taken by the camera, looking at a heating source with a current of 0.15 A being driven. The difference between the camera and the thermocouple were much smaller, as seen below. I'd still like to take a few more measurements to solidify that this has rectified the issue, but at the moment it seems this is working much better. Images at: https://drive.google.com/drive/folders/1VDvZ1rfEGWsHq1AgG-chde3Cd8Do_piD?usp=sharing
Attachment 1: Screenshot_2023-10-23_at_11.06.04_AM.png
Screenshot_2023-10-23_at_11.06.04_AM.png
  249   Mon Oct 30 11:03:56 2023 TylerUpdateScripts/Programs 

FLIR: After some adjustments, the plot generated from the FLIR measurements look much more symmetric (see attachment). There are more included grid points, which smooths out the curve as compared to last week.

Red Pitaya: It looks like a new OS update was released for the RP, which includes a new Python API (was previously only available in C). I'm going to try and update the one we have currently running in lab.

Attachment 1: Screenshot_2023-10-23_at_12.39.05_PM.png
Screenshot_2023-10-23_at_12.39.05_PM.png
  255   Mon Nov 6 11:29:24 2023 TylerUpdateElectronicsPreliminary RTD Calculations, RP Software Update

Preliminary RTD calculations are shown below, given an input of 10 V and desiring a few mA of current. It looks like R_ref should be at least 1 kOhm (refer to plots/circuit below), keeping in mind we need to have <10 V input for the ADC.

RP: The Red Pitaya Software was updated to OS 2.00. All examples on the RP website should run without issue.

Attachment 1: cvt.png
cvt.png
Attachment 2: VvC.png
VvC.png
Attachment 3: IMG_7471.jpg
IMG_7471.jpg
  264   Mon Nov 13 11:07:50 2023 TylerUpdateVACRTD Analysis

After initial analysis from last week on a single RTD, I then extended to looking at all 8 in series with R_ref (set to 1 kOhm). Shown below are the edge cases for the setup:

  • RTDs are all at ambient lab temperature. This would correspond to a minimum resistance value.
  • RTDs all read out 400 C. This gives the maximum resistance value.

The results show that indeed only a few mA of current is drawn even at room temperature (a little above 5.5 mA), and this will continue to decrease with increasing temperature. The voltage across a single terminal, at a maximum, is only about 5.4 V.

Attachment 1: Screenshot_2023-11-13_at_11.05.45_AM.png
Screenshot_2023-11-13_at_11.05.45_AM.png
Attachment 2: Screenshot_2023-11-13_at_11.06.04_AM.png
Screenshot_2023-11-13_at_11.06.04_AM.png
ELOG V3.1.3-7933898