ligo-ex ligo-ds
  Richardson Lab Experimental Log, Page 7 of 14  Not logged in ELOG logo
ID Date Author Type Category Subjectdown
  431   Thu Aug 29 16:39:34 2024 LukeUpdateVACPhotos of heater tape
Attachment 1: IMG_2052.jpeg
IMG_2052.jpeg
Attachment 2: IMG_2051.jpeg
IMG_2051.jpeg
Attachment 3: IMG_2048.jpeg
IMG_2048.jpeg
Attachment 4: IMG_2047.jpeg
IMG_2047.jpeg
Attachment 5: IMG_2046.jpeg
IMG_2046.jpeg
Attachment 6: IMG_2044.jpeg
IMG_2044.jpeg
  560   Tue Apr 22 13:00:04 2025 NathanUpdateVLC UpdatePhotodiode signal check and additional signal check due to 20Mhz signal and amplifier
In general, photodiode was able to achieve signal. Below are the voltage signals and frequency measurements of the photodiode with the dc bias and amplifier connected to the EOM (Vpp = 2mV for first and Vpp= 1 for second). Also I need assistance in using the PZT driver.
Attachment 1: thumbnail_(8)_(1).jpg
thumbnail_(8)_(1).jpg
Attachment 2: vpp.jpg
vpp.jpg
  614   Tue Jul 15 13:13:48 2025 ChristinaUpdateTCSPhotodetector optical post assembly
Attachment 1: pdreassembly.png
pdreassembly.png
  94   Tue May 9 15:49:20 2023 JonInfrastructureVACPermanent cable routing

Today I brought in a fresh supply of zip ties (we now have 1500 in the tool chest) and used them to permanentize the cable routing for the gauges, pumps, and RGA.

I also brought and installed a 3-foot 15A extension cable for powering the scroll pump. Installing the cable required shutting down the pumps, which I did and then reverted via the following procedure:

  1. Close the 4.5" gate valve, 2.75" gate valve, and the bypass line angle valve.
  2. Shut down the turbo pump.
  3. Shut down the scroll pump.
  4. Unplug the scroll pump and install the extension cable.
  5. Power on the scroll pump.
  6. Power on the turbo pump.
  7. Open all three valves.

Incidentally, before I started, I noticed that the pressure in the main volume had reached 7E-7 torr, which is lower than the pressures seen last week. The system quickly returned to this pressure after I restarted the pumps.

  12   Mon Nov 28 18:10:23 2022 shaneUpdateELOGParticle counts in the clean room
Particle count stats for the clean room Nov 28, 2022: Took 10 sample runs in each of 5 regions in the clean room (5 runs per region with a person inside the clean room for the measurement, and 5 runs per region without anyone in the clean room for the measurement), for a total of 50 samples taken. Sample time was 60 seconds. Overall clean room average particle count for the size ranges are as follows: 0.3 micrometers- 3405.76 (room occupied), 974.92 (room empty) 0.5 micrometers- 409.72 (room occupied), 409.72 (room empty) 1.0 micrometers- 1102.2 (room occupied), 282.6 (room empty) 2.5 micrometers- 692.32 (room occupied), 183.68 (room empty) 4.0 micrometers- 254.28 (room occupied), 84.72 (room empty) 5.0 micrometers- 141.24 (room occupied), 84.72 (room empty) 7.0 micrometers- 56.48 (room occupied), 84.72 (room empty) 10.0 micrometers- 42.36 (room occupied, 42.36 (room empty) More statistics (including individual stats on the 5 regions within the clean room) attached.
Attachment 1: clean_room_particle_counts_11_28_-_Sheet1.pdf
clean_room_particle_counts_11_28_-_Sheet1.pdf
  64   Tue Apr 4 16:52:08 2023 JonUpdateCleanroomParticle counter docking station
Due to the problems we have encountered trying to serially communicate directly with the Met One DR-528 particle counter, I went ahead and ordered the add-on communications/charging dock. The dock "ethernetizes" the serial communications by wrapping them in TCP packets and transmitting them over the lab network. This will allow our Python code to communicate using the standard Internet sockets package, rather than PySerial, PyNut, etc. the docking station arrived today and I delivered it to the lab. I left it laying next to the particle counter in the electronics rack.
Attachment 1: Met_One_Dock.png
Met_One_Dock.png
  71   Wed Apr 12 16:17:50 2023 shaneUpdateGeneralParticle Counter moved
With the added height of the new docking station, the particle counter no longer fits under the Windows monitor in the electronics rack and has been moved to the desk in the corner of the lab.
Attachment 1: IMG_6693.HEIC
  235   Mon Oct 9 11:29:37 2023 TylerUpdateElectronicsPSD/CSD Plot Updates
Continuing from Sophia's SURF Project this summer: Plotting the Power-Spectral Densities (PSDs) and Cross-Spectral Densities of two signals. A continued issue that has been observed in the plots is the random "jump" of one PSD curves. Below, a 1.4 kHz signal is driven with a magnitude of 0.7 Vpp, and connected to the two output ports of the Red Pitaya via an SMC T-Adapter. At the moment, I anticipate that this bug might have something to do with the Real-Time GUI code being used, since this hasn't been observed when running the calculations without it (see below).
Attachment 1: Screenshot_2023-10-09_at_11.20.25_AM.png
Screenshot_2023-10-09_at_11.20.25_AM.png
Attachment 2: welch_csd-2.png
welch_csd-2.png
  536   Mon Mar 10 21:13:12 2025 NathanUpdateVLC UpdatePDH Lock Progress and Next Steps
Progress update: Photodiode is set up, need to set up DC bias tee, EOM has 400-600nm wavelength range with a resonance frequency of 0.1-100MHz. Additionally need to connect the DC bias to SR560, put it in low pass, then connect to the function generator and connect it to oscilloscope to observe. Once everything looks good and working, need to connect the driver and we can start actuating on the cavity ETM.
Attachment 1: thumbnail_(2).jfif
Attachment 2: thumbnail_(4).jfif
Attachment 3: thumbnail_(5).jfif
Attachment 4: Visible_Light_Optical_Cavity_(2).pdf
Visible_Light_Optical_Cavity_(2).pdf
  551   Tue Apr 8 20:04:35 2025 NathanUpdateVLC UpdatePDH Lock Progress
Connected the DC bias and mixer. Need to connect signal generator now to RF Amp and mixer, and also need to connect power supply. After that will see if I can get a detection on the oscilloscope and if so will set up the driver. [Photos attached]
Attachment 1: IMG_5450.jpeg
IMG_5450.jpeg
Attachment 2: IMG_5449.jpeg
IMG_5449.jpeg
  39   Wed Mar 1 21:27:20 2023 JonInfrastructureCleanroomOver-table shelf height raised

Jon, Cao

Today we raised the height of the shelf overhanging the cleanroom laser table by 8 inches. This was done to create more vertical clearance between the top-loading vacuum chamber and the bottom of the shelf. The added clearance should make both removing the chamber lid and inserting large parts easier.

The procedure required unmounting the shelf and removing all eight vertical support posts (1" x 1" x 18.5" pieces of 80/20 unistrut). The support posts were taken to the machine shop and cut, retapped, and cleaned (coarsely, with IPA wipes) prior to reinstallation. We took care to minimize the contamination introduced into the cleanroom, but some amount of particulate from disturbing the shelf was unavoidable.

This work is completed, and the cleanroom is now ready for final cleaning (HEPA vac, mopping, and wiping down of all surfaces including the softwalls).

Attachment 1: cleanroom-shelf.jpg
cleanroom-shelf.jpg
  16   Mon Jan 23 17:26:15 2023 Peter CarneyUpdateGeneralOven cleaning
Aiden and Cao Turned on the oven to 120 deg C for 12 hours. After 12 hours, put it at 200 deg C for 48 hours.
  17   Thu Jan 26 18:58:25 2023 AidenUpdateGeneralOven Heating
Aiden cleaned the oven with methanol again and set the oven to 260 degC for 12 hours.
  338   Fri Feb 23 18:03:27 2024 TylerUpdateFLIROptical Test Setup in Cleanroom
[Tyler, Xuesi]

The FLIR and test mass stand-in have been transferred into the cleanroom. A software test will be run as soon as we get an ethernet cable long enough to reach into the cleanroom where the camera is set up. Once this is finished, the FLIR will be moved aside for construction of the FROSTI! When completed, the camera will be placed back into position for in-air optical testing.

Attachment 1: IMG_0941.png
IMG_0941.png
  9   Tue Jul 26 14:10:35 2022 CassidyUpdateCamerasOptical Post Replacement and Realignment
Today I replaced the 3" optical post that the camera rests on with a 2" optical post in order for the screen to correctly fill out the camera's FOV. The 3" post is now in the glass optics cabinet next to the FLIR camera configuration box, wrapped in the protective materials from the 2" post for safekeeping.

There were no issues with the physical replacement of the post, except that the fork clamp on the post needs to be on one of the perpendicular, not diagonal, axes in order to be secure. I chose the front axis (towards the screen) as before in order to easily access the alignment knobs.

To align, I followed the same process as last time except for a more purposeful original rough alignment. For alignment purposes, the visual camera was used. For the first rough alignment, I pulled the camera as far back as possible on the x axis (with the z axis roughly centered), then moved the entire stage setup back until I could just see both the top and bottom edge of the screen. Then, I set the z-axis to an extreme in order to use the edge of the screen to align the rotational and y-axis pieces for the fine alignment.

For the fine alignment, starting with the z and x axis at extremes, I began by aligning the rotational axis. To do this, I used the gaps between the top of the screen and the camera window on the far left and right of the image. When these gaps were equal I knew the rotation was adequately set. Then, I set the y-axis so that the pattern was centered. If the gaps were no longer even, I redid the rotation alignment and ditto with the y axis until both were set. This resulted in a rotation of about two degrees and a y axis at just under 3.5.

To set the z and x axis, I centered the z-axis using the top and bottom of the screen, which should both be visible if the rough alignment was done correctly. Then, I adjusted the x axis by pushing it forward until the top and bottom of the screen were just out of the frame. As with the rotational and y-axis, I iteratively fine tuned the x and z axis until both the image was centered in the z axis and only the screen was in view. This resulted in a z-axis value of just over 5.5 and an x axis value of nearly 1.75.

Pictures are included of all alignment knobs and the new post/stage setup!

Attachment 1: YAxisAlignment.png
YAxisAlignment.png
Attachment 2: XRotZAlignment.png
XRotZAlignment.png
Attachment 3: NewOpticalPost.png
NewOpticalPost.png
  527   Thu Feb 27 18:22:21 2025 ShaneUpdateCDSOngoing work in server rack and desk area (1119)
As part of the vacuum communications debugging process, there is ongoing work in the server rack and desk area in 1119. Please be careful not to move any of the materials left out in the server rack or on the work bench.
  650   Tue Sep 30 12:51:23 2025 TylerUpdateTCSO6 RH Update 09/30/2025
Slides
  636   Tue Aug 26 13:09:03 2025 TylerUpdateTCSO6 RH Update
[Tyler]

Below are slides giving an update on RH O6 design progress.

Slides
  424   Mon Aug 12 17:09:04 2024 Cynthia UpdateInterferometer SimulationsO4 vs O5 cavity scans with ITM plume guess or ETM
So far the proposed solution for what to use for O5 ITM mirror for SIS simulations is either the plume guess file or using the O5 ETM mirror for ITM. I had ran a cavity scan for both of these possibilities and it seems like the results produced are fairly different. The next step is most likely moving on to adding the Frosti effects including both of the proposed possibilities, and before this, a few confirmations and verifications are needed using similar previous studies to make sure my code produce the right result.
Attachment 1: O5_plots_and_tables.pdf
O5_plots_and_tables.pdf
Attachment 2: O4_plots_and_tables.pdf
O4_plots_and_tables.pdf
Attachment 3: ETM12mapO5.png
ETM12mapO5.png
Attachment 4: Screenshot_2024-08-12_at_5.05.31_PM.png
Screenshot_2024-08-12_at_5.05.31_PM.png
  392   Mon Jul 1 15:19:42 2024 Cynthia UpdateInterferometer SimulationsO4 and O5 mirror cavity scans with thermal effect
Attachment 1: cavityScan_07_01.pdf
cavityScan_07_01.pdf
  578   Thu May 15 12:03:41 2025 Xuesi MaInfrastructureElectronicsNotice

All Sorensen power supplies are turned off. The Cymac system is down (All Chassis are down)

If you need more information or if you need to turn them back on please contact Xuesi Ma.

  67   Tue Apr 11 13:15:40 2023 CaoInfrastructureClean & BakeNitrogen gas tank replaced
The nitrogen gas tank has been replaced with a new unit. The new tank is ultra pure nitrogen (>99.9% nitrogen). The new tank has been placed and secured to the rack where the old one is; cap is removed and regulator is reinstalled onto the new tank.
  26   Fri Feb 10 16:34:45 2023 Huy Tuong CaoInfrastructureClean & BakeNitrogen gas tank ready to use
Cao,

Today I fixed the final bit related to the nitrogen gas tank, which is to apply sealing tape to M-NPT connector of the hose to prevent leakage (file: AirGunSealed.jpg)
After application of the tape, no audible leak can be heard from connection between the hose and the air gun.

The general operating procedure for the gas tank is as following:

  1. Turn the regulator (blue handle) anti-clockwise still it's loose
  2. Turn the valve on nitrogen as tank anti-clockwise, immediately the RHS meter of the regulator would jump to approx 2000 psi. This is the standard pressure for high pressure gas tank
  3. Turn the regulator clock-wise slowly until the pressure one the LHS meter face reads approx 60 psi. This is sufficient for drying parts with. At this point, the flow pressure still should register zero
  4. Press the trigger on the air gun, a high pressure air flow should come out and the flow meter should increase
  5. When finished, close the gas tank valve, turn the regulator anti-clockwise, then press the air gun trigger to release gas left in the hose/gun

Attachment 1: AirGunSealed.jpg
AirGunSealed.jpg
  412   Fri Jul 26 17:25:42 2024 JonInfrastructureComputersNew workstation ws4 installed in 1129

Today I installed the second desktop workstation in 1129. The new machine is an Intel NUC13ANHi5, with a 12-Core Intel i5-1340P CPU, 32GB DDR4 RAM, and a 1TB SSD.

I loaded it with a fresh installation of Debian 12 and installed the LIGO CDS workstation (control room) tools. It is assigned the hostname ws4 and and the static IP address 192.168.1.19 on the local lab network. Like the other CDS workstations, there is just one user account accessible with the usual credentials.

The machine is fully set up and ready for use.

Attachment 1: ws4.jpg
ws4.jpg
  379   Thu Jun 20 18:01:11 2024 JonInfrastructureComputersNew workstation ws3 in 1129
I have repurposed a spare Intel NUC to serve as the first of two planned workstations in 1129. Today I set it up with a fresh installation of Debian 12.5, assigned it the static IP address 192.168.1.15, and installed the LIGO CDS workstation packages. It is assigned the host name ws3. Temporarily, the mouse is borrowed from ws1 in 1119. We need to order another mouse as well as a 25-ft Ethernet cable for connecting the workstation to the 1129 switch, via the ceiling.
  385   Wed Jun 26 15:33:41 2024 JonInfrastructureComputersNew workstation ws3 in 1129

Set-up of the first CDS workstation for 1129, ws3, is complete and the machine is ready for use. The login credentials are the same as the other lab machines.

All that now remains is to install a permanent cable tray for running the new Ethernet cables between the electronics rack and bench (they are currently dangling from the suspended lights).

Quote:
I have repurposed a spare Intel NUC to serve as the first of two planned workstations in 1129. Today I set it up with a fresh installation of Debian 12.5, assigned it the static IP address 192.168.1.15, and installed the LIGO CDS workstation packages. It is assigned the host name ws3. Temporarily, the mouse is borrowed from ws1 in 1119. We need to order another mouse as well as a 25-ft Ethernet cable for connecting the workstation to the 1129 switch, via the ceiling.

Attachment 1: IMG_1368.jpg
IMG_1368.jpg
  198   Wed Aug 9 11:32:56 2023 JonInfrastructureGeneralNew power cables

I delivered new NEMA 5-15 (120V / 15 A) power cables to the lab for the following items:

  • WS2 (cleanroom) cart - 10ft cable
  • Electronics workbench overhead LED - 10ft cable
  • Both PI heater controller sets - (2) 6ft cables

I installed the new cables on the WS2 cart and the workbench myself, and left the two 6ft cables (as pictured below) for Aiden to install on the PID controllers after the current bake is finished.

Attachment 1: IMG_0299.jpg
IMG_0299.jpg
  562   Fri Apr 25 19:18:10 2025 Luke UpdateScripts/ProgramsNew optimization plots

I have created two more plots. One on the transmitted power for a specific configuration. This was done by taking the product of the efficiency and the power released by surface area of the source plane for a configuration assuming it is at 300C. The second is a plot of how close the reflector is to the test mass I have included a contour line at 2mm the nominal proximity constraint. Both of these plots are generated for a 2cm wide target plane.

Somethings of note: Because the proximity requires no modeling and can be determined purely mathematically it is easy to re-generate this plot of other target widths. If you increase the target width the proximity contour moves upward and downwards if reduced.

Attachment 1: effective_power.png
effective_power.png
Attachment 2: proximity.png
proximity.png
  575   Sun May 11 19:36:42 2025 Luke UpdateScripts/ProgramsNew optimization plots

Slides

I have re-run the optimization plots with the changes mentioned last week. 

I have added the heat maps of the HOM scattering from HR deformation and the HOM scattering due to the OPD difference as well as the curvature actuation of the surface. 

  640   Tue Sep 2 18:33:38 2025 LiuInfrastructureGeneralNew lab room 1125 layout and measurement

Action items for facilities

 

  • Ceiling lights replacement
  • Fume hood cleaned and refurbished
  • Sink countertop cleaning
  • Floor cleaned and waxed
  •  
Attachment 1: 1125.pdf
1125.pdf
  135   Tue Jun 27 15:18:26 2023 PamellaUpdateFLIRNew data collection
    [Pamella]
    Yesterday, I did a new data collection. I could get a better data in this time and I realized than I need use some angle on the reflector for got a better shot. I need do that because the FLIR camera just is able to get the triangle shape for complete if I keep exactly in same line for center point on the camera but for the data now I need move the refletor to go up and go down. So wasn't working very well if I keep the reflector without angle.
    Now I am using the refletor with a angle. In the more high part I turn the mask for look the table and in the lower part I turn the mask to look up. I attached a photo below for this configuration. Also the processes to get the data was the same than I using last time, the only diference is now I just moved the pillar after get data for high position and lower position for the same reference point on the triangle shape.
    Note: In most data it is impossible to keep the same parameter for current, voltage and temperature. Most have a small variation but not a big difference. For example: I got the temperature in the first position 46.8 °C and in the second position I got 47.1 °C, it's just an example. It's just so we know that we don't have the exact same parameters all the time. In my opinion this is not a problem because it is just a small variation.
    The process to get data
  • 09:35 am: Turned on the device (current on). I wanted for 30 minutes before start get data.
  • 10:05 am: Started taking snap on position one (Reference point: -0.10, 0.050). Parameters:0.11A,1.7V,46.8°C
  • I took four snap on the same position for compare after on data analyzes. I just wanted one minute break between the snaps. I did the same for every position.
  • 10:15 am: Started taking snap on position two (Reference point:-0.10,-0.050 ). Parameters: 0.11A,1.7V,47.1° C
  • 10:22 am: Started taking snap on position three (Reference point:0.00 ,0.050). Parameters: 0.11A,1.6V,48.1°C
  • 10:26 am: Started taking snap on position four (Reference point: 0.00,-0.050). Parameters:0.11A,1.6V,48 °C
  • 10:32 am: Started taking snap on position five (Reference point: 0.05,-0.050). Parameters:0.11A, 1.6V, 48°C
  • 10:37 am: Started taking snap on position six (Reference point: 0.05,0.050). Parameters: 0.11A,1.7V,47.6 °C
  • We can see in the photos attached below every position also I am working in the analyzes.
Attachment 1: FLIR_data.002.jpeg
FLIR_data.002.jpeg
Attachment 2: FLIR_data_2.002.jpeg
FLIR_data_2.002.jpeg
  123   Tue Jun 13 21:33:46 2023 PamellaUpdateGeneralNew cabinet for PPI
[Pamella and Dr.Richardson]
    Dr. Richardson installed the new cabinet outside the room for the PPI equipments, and then I wiped all surfaces inside and outside of the new cabinet.
    Dr.Richardson and I have finished organizing the PPI within the new cabinet. Also we have new supplies of gloves, and will likely have more supplies for other PPIs soon.
  157   Fri Jul 14 16:49:42 2023 PamellaUpdateFLIRNew bridge and calibration data
[Pamella]
    Aiden 3D printed a new bridge for the heater and I installed the new bridge yesterday.
    I started collecting data to plot a calibration with the heater. I'm doing measurements with current and the thermocouple (thermometer) to compare with FLIR measurements and have a good calibration.
Attachment 1: IMG_8606.jpeg
IMG_8606.jpeg
  367   Mon May 13 14:48:27 2024 AidenUpdateVACNew Vacuum Chamber Design
I created a new model for the vacuum chamber based on the ideas from last week's meeting.

It should be noted that we can not have the zero length reducer and the gate valve on the same arm as they both have blind tapped holes. We need to decide if we want the gate valve or not as if we do, we will need to use the 3 inch long reducer.

Important Measurements:

From the back of the flange to the back of the RGA: 25.25 in

Height from bottom of the blank to the bottom of the chamber: 2.75 in

Attachment 1: VaccumChamberAssembly__Redesign_v5.jpg
VaccumChamberAssembly__Redesign_v5.jpg
  398   Mon Jul 8 17:01:45 2024 ShaneUpdateCDSNew CyMAC internal layout
[Jon, Shane] Internal layout of CyMAC has been updated (labeled image attached) to accommodate the replacement of the two ribbon cables. Looking down on the chassis from the front, and going from left to right, the new placement is as follows: BIO card, DAC adapter board, ADC adapter board, DAC card, ADC card. NOTE: As part of ADC-DAC loopback testing, we're disconnecting from the FROSTI readout chassis and using the cables to connect directly from ADC input channels to DAC output channels . Initial testing confirmed functionalility of all 32 ADC input channels and the first 8 DAC output channels.
Attachment 1: updated_cymac.jpeg
updated_cymac.jpeg
  108   Tue May 30 14:51:27 2023 PamellaUpdateFLIRNew 3D mask - Bigger triangles shapes.
[Pamella]
    Change the 3D mask
  • Started removing the old mask in the reflector.
  • Started cutting the new aluminum lid for the new 3D mask and assembling in the 3D mask.
  • Started doing the new cable connections for the light.
  • Started testing the resistance in the new cable connections. The multimeter show us 14.2 Ω.
  • Started attaching the mask with the light and the thermometer sensor.
  • Started assembling the 3D mask in the reflector.
  • Started the initial tests.
  • 2:00 pm: Started testing the new mask. Parameters: 1.4 V 0.1A and 27 C
  • 2:38 pm: Started taking snap. Final parameters:1.4 V 0.1A and Temperature: 39.9 C
  • Started doing some adjustments in the triangles shapes for the futures tests.
Attachment 1: AcquisitionImage(May-30-2023_14_38).jpg
AcquisitionImage(May-30-2023_14_38).jpg
Attachment 2: thumbnail_IMG-7303.jpg
thumbnail_IMG-7303.jpg
  643   Tue Sep 9 12:13:38 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI optimization for A# with grid search
Multi-ring FROSTI optimization based on IFO metrics for A# with grid search
Attachment 1: Update.pdf
Update.pdf
  616   Tue Jul 29 11:27:57 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI optimization for A#
Attachment 1: Update.pdf
Update.pdf
Attachment 2: scene-export-temp.html
Attachment 3: scene-export-deformation.html
  625   Tue Aug 12 12:16:21 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI optimization for A#
Multi-ring FROSTI optimization for A# with static polishing
Attachment 1: Update.pdf
Update.pdf
  647   Tue Sep 23 12:39:29 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI optimization for A#
Attachment 1: Update.pdf
Update.pdf
  652   Thu Oct 9 12:18:14 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI optimization for A#
Attachment 1: Update.pdf
Update.pdf
  654   Thu Oct 16 10:31:20 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI optimization for A#
Two-ring FROSTI grid search with 4D parameter space search and visualization
Attachment 1: interactive_plot_dropdown.html
Attachment 2: Update.pdf
Update.pdf
  655   Thu Oct 23 10:07:51 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI optimization for A#
4D parameter space visualization and dimensionality reduction for better optimization
Attachment 1: loss_embedding.html
Attachment 2: Update.pdf
Update.pdf
  658   Thu Oct 30 11:45:44 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI optimization for A#
Multi-ring FROSTI optimization for A#: dimensionality reduction with PCA vs. t-SNE
Attachment 1: Update.pdf
Update.pdf
  633   Tue Aug 26 12:14:24 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI design for A# with ring heater optimization
Multi-ring FROSTI design for A# with ring heater optimization
Attachment 1: Update_(1).pdf
Update_(1).pdf
  639   Tue Sep 2 12:41:33 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI design for A# with ring heater optimization
[Tyler, Liu] Ring heater optimization, continued
Attachment 1: Update.pdf
Update.pdf
  172   Tue Jul 25 15:35:14 2023 ShaneUpdateElectronicsMore chassis moved
Finished basic assembly of binary input and output chassis today, and moved one to top shelf of work bench. The other is in corner of work bench (image attached). They are delicate, so please do not move them or place anything on top of them.
Attachment 1: IMG_8409.HEIC
  91   Wed May 3 19:03:47 2023 Julian,Pamella and CaoPhysicsVACModify vacuum asssembly and install RGA
[Pamella, Julian, Cao]
    Today we started re-configuring the vacuum chamber components.
    Particle count
  • 10:45 am : Starting the particle count in the clean room.
  • 11:13 am : Finished the particle count in the clean room.
  • Zone 3 :
    • 0.3 u: 2535
    • 0.5 u: 1413
    • 1.0 u: 789
  • Zone 4 :
    • 0.3 u: 457
    • 0.5 u: 290
    • 1.0 u: 207
      Starting the reconfiguration.
    • 11:21 am: Started removing up-to-air valve
    • 11:30 am: Finished removing up-to-air valve.
    • 11:33 am: Started removing calibrated Ar Leak and the elbow.
    • 11:35 am: Finished removing calibrated Ar Leak and started assembling the up-to-air valve.
    • 11:47 am: Finished assembling up-to-air valve.
    • 11:48 am: Started removing magnetron gauge.
    • 11:55 am: Finished removing magnetron gauge.
    • 11:57 am: Started installing the elbow and calibrated Ar Leak and started removing the RGA probe
    • 12:13 pm: Finished installing the elbow and calibrated Ar Leak.
    • 12:20 pm: Break for lunch.
    • 01:25 pm: Come back from lunch break.
    • 01:30 pm: Started removing RGA line.
    • 01:39 pm: Finished removing RGA line.
    • 01:39 pm: Started removing gate valve [RGA line] and finished removing gate valve.
    • 02:00 pm: Started installing gate valve in the new position.
    • 02:08 pm: Fished installing gate valve [RGA line].
    • 02:09 pm: Checked the screws in the elbow to gate valve. [RGA line].
    • 02:28 pm: Finished checking the screws in the elbow to gate valve [RGA line].
    • 02:29 pm: Started installing RGA line.
    • 02:39 pm: Finished installing RGA line.
    • 02:50 pm: Started installing magnetron gauge.(In this part we assembled gauge with the used gasket)
    • 03:02 pm: Finished installing magnetron gauge.
    • 03:04 pm: Started installing RGA probe [RGA line].
    • 03:13 pm: Finished installing RGA probe [RGA line].
    • 03:15 pm: Connected the cables to the magnetron gauge.
    • 03:20 pm: Started the testing in the vacuum chamber.
    • 04:30 pm: Started installing the RGA 200 and connected the cables (power cable and DB9 cable).
    • 04:52 pm: Finished installing the RGA 200 and connected the cables (power cable and DB9 cable).
    • 04:53 pm: Started testing the RGA connections. Logrus is able to recognize and connect to RGA unit. We are leaving the turbo pump on for a few hours before checking back for pressure readouts.
    • 05:00 pm : Starting the particle count in the clean room.
      1. Zone 3 :
        • 0.3 u: 2244
        • 0.5 u: 997
        • 1.0 u: 207
      2. Zone 4 :
        • 0.3 u: 2993
        • 0.5 u: 1579
        • 1.0 u: 498
    • 05:30 pm : Finished the particle count in the clean room.
    • 05:34 pm: Test pressure readout from the gauges.
      1. Channel 1 : 1.06E-5 mbar
      2. Channel 2 :8.89E-6 mbar
      3. Channel 3 :5.0E-4 mbar
    • 05:59 pm: Test pressure readout from the gauges.
      1. Channel 1: 9.09E-6 mbar
      2. Channel 2: 7.75E-6 mbar
      3. Channel 3: 5.0E-4 mbar
Attachment 1: vaccum_chamber_re-configuration.jpg
vaccum_chamber_re-configuration.jpg
  111   Wed May 31 16:18:41 2023 AidenSummaryGeneralMock FROSTI
Finished assembly on the Mock FROSTI so it is ready for the fit test at LIGO lab.
Attachment 1: IMG_4394.jpg
IMG_4394.jpg
  397   Mon Jul 8 12:26:10 2024 MichaelUpdateInterferometer SimulationsMinimizing Astigmatic Effects in the SRC of LIGO
This is the first look and analysis of the effects of astigmatism in the aLIGO optical layout which can lead to mode mismatch and therefore increased losses into higher order modes. Specifically I am currently looking at measuring the effect of changing the radii of curvature of the SRC mirrors https://docs.google.com/presentation/d/1qhIehqyNukg4g8S2fqfQZz83yr9aB76tEa6fn2J-TrM/edit?usp=sharing
ELOG V3.1.3-7933898