Tue Apr 22 12:08:46 2025, Xuesi Ma, Update, ,  
|
Test 7: Reconnect ribbon cables
- Setup: Reconnect all the ribbon cables, and turned on the AI Chassis while isolated rack
- Observation: No spike observed.
Test 8: Reconnect to rack
- Setup: Reconnect the AI Chassis to the rack.
- Observation: No spike observed.
Test 9: Restore to original position.
- Setup: CLose up the AI Chassis, and put it back to its original location.
-
- Observation: No spike observed.
|
Tue May 6 13:19:31 2025, Xuesi Ma, Update, , 
|
| |
Tue May 13 12:35:01 2025, Xuesi Ma, Update, ,  
|
| |
Mon May 19 10:59:14 2025, Xuesi Ma, Infrastructure, ,
|
Status Update
Previous Status:
All Sorensen power supplies were turned off. The Cymac system was down (all chassis were offline).
If you need more information or assistance in turning them back on, please contact Xuesi Ma.
Update: The systems are now back online.
Note: The Sorensen power supplies’ voltage can now be controlled via channel VEXC8 of the DAC.
If scripting is required, the ezca Python package is highly recommended for interfacing and control.
|
Tue May 27 12:52:22 2025, Xuesi Ma, Update, , 6x
|
Using the configuration in section 2.11.1 in the Sorensen installation and operation manual (attached below), I'm able to remote control the Sorensen power supply.
I use DAC VEXC8 as the external voltage source and a breakout board to match the configuration of connector J3.
I have recorded 7 cycles (attached below), and am now working on analyzing the data. |
Tue Jun 3 13:08:47 2025, Xuesi Ma, Update, , 
|
We currently have the parts for a DB9 connector but need parts for a DB25 connector, where did we get the parts for the DB9 connector? We may be able to use the same company for the DB25 connector.
Also need a cable for the connectors. Find a DB9 to DB9 cable in 1129, not sure if I can disassemble it and use just the cable.
The pressure in the vacuum chamber rises to 2.12e-8 torr, which needs baking. |
Tue Jun 3 13:16:53 2025, Xuesi Ma, Infrastructure, ,    
|
|
Date and Time: Around 3:50 PM on June 2, 2025
Location and Temperature:
- Back of the room 1119, around the working station: 86 °F
- Front of the room 1119, around the doorway: 84.1 °F
- Back of the room 1129, around the working station: 78.6 °F
- Front of the room 1129, around the doorway: 78.3 °F
- In the hallway in front of room 1119: 75.2 °F
| |