ligo-ex ligo-ds
  Richardson Lab Experimental Log, Page 14 of 14  Not logged in ELOG logo
ID Date Author Type Category Subjectdown
  594   Tue Jun 17 12:22:43 2025 Xuesi MaUpdate  
CYMAC Remote Control Cable Documentation

CYMAC Remote Control Cable Assembly

Cable Functions

The custom cable assembly serves two primary functions:

1. Jumper Connections (DB25 Side - Connector J3)

The following pins are jumpered on the DB25 connector:

From Pin To Pin
Pin 3 Pin 16
Pin 10 Pin 11
Pin 12 Pin 24
Pin 13 Pin 25
Pin 22 Pin 23

2. Signal Wiring (DB9 to DB25)

The following connections are made between the DB9 and DB25 connectors:

DB9 Pin DB25 Pin Function
Pin 1 Pin 9 Positive side
Pin 6 Pin 12 Negative side
Pin 5 Pin 6 Ground

Note: All connections are verified with a multimeter.

Note: This cable allows remote control of the Sorensen power supply with an external voltage source (Cymac DAC).

Note: For more details, check section 2.11.1 from the attached Sorensen power supply manual.

Attachment 1: 20250616_132743.jpg
20250616_132743.jpg
Attachment 2: 20250616_132945.jpg
20250616_132945.jpg
Attachment 3: 20250616_133235.jpg
20250616_133235.jpg
Attachment 4: 20250616_180103.jpg
20250616_180103.jpg
Attachment 5: 20250616_180129.jpg
20250616_180129.jpg
Attachment 6: 20250616_180132.jpg
20250616_180132.jpg
Attachment 7: 20250616_180317.jpg
20250616_180317.jpg
Attachment 8: 20250616_181011.jpg
20250616_181011.jpg
Attachment 9: 20250616_181033.jpg
20250616_181033.jpg
Attachment 10: 20250616_181052.jpg
20250616_181052.jpg
Attachment 11: 20250616_132759.jpg
20250616_132759.jpg
Attachment 12: operating-manual-Sorensen-DCS-programmable-switching-dc-power-supplies-1kw-dcs-e.pdf
operating-manual-Sorensen-DCS-programmable-switching-dc-power-supplies-1kw-dcs-e.pdf
  596   Tue Jun 24 11:50:15 2025 Xuesi MaUpdate  
Heater Element Resistance Test Results

Heater Element Resistance Test Report

Test Overview

Conducted several on-and-off tests on heater elements at 24V, followed by switching heater elements 5 and 8 by changing their connection pins to the ADC.

Resistance Measurements (Ω)

Heater Element Before Switch After Switch Ladder Test Difference (After-Before)
Element 1 153.09577918 153.11447008 153.83373894 +0.0186909
Element 2 149.70967557 149.8175605 150.08915205 +0.10788493
Element 3 157.88314236 157.87136021 157.39696467 -0.01178215
Element 4 146.80153622 146.78826747 147.77923177 -0.01326875
Element 5 151.80888657 152.2204132 150.44919044 +0.41152663
Element 6 148.24375881 148.32846152 146.89161192 +0.08470271
Element 7 144.72286546 144.76879963 145.03627875 +0.04593417
Element 8 151.4564267 149.14240417 149.88675562 -2.31402253

Supporting Graphs

Attached below are the resistance-time graphs for all heater elements:

  • Complete test duration graph
  • Pre-switch period graph
  • Post-switch period graph
  • Resistance voltage graph after removing data from after power outage

Analysis & Conclusion

The switch did not physically change the resistance of the heater elements, suggesting the measurements reflect their true resistances.

Attachment 1: resistance_vs_time_all_channels_channel_switch.jpg
resistance_vs_time_all_channels_channel_switch.jpg
Attachment 2: resistance_vs_time_all_channels_channel_switch_one_cycle_before.jpg
resistance_vs_time_all_channels_channel_switch_one_cycle_before.jpg
Attachment 3: resistance_vs_time_all_channels_channel_switch_one_cycle_after.jpg
resistance_vs_time_all_channels_channel_switch_one_cycle_after.jpg
Attachment 4: resistance_vs_voltage.jpg
resistance_vs_voltage.jpg
  603   Tue Jul 1 13:41:36 2025 Xuesi MaUpdate  
[Ma, Liu]

Slides

  608   Thu Jul 10 16:41:18 2025 ChristinaUpdateTCS 

[Christina, Ma, Tyler]

Pomona Box completed and is now on the server rack in 1119. Additionally, the photodetectors were connected to the box and set up in front of the heater elements.

Attachment 1: 2994921113032912040.JPG
2994921113032912040.JPG
Attachment 2: 4717009101108174053.JPG
4717009101108174053.JPG
Attachment 3: 5782646701589802125.JPG
5782646701589802125.JPG
  611   Tue Jul 15 10:40:23 2025 ChristinaUpdateTCS 

[Christina, Ma, Tyler]

Yesterday, Ma and I started taking data for the PSD/CSD measurements while the photodetectors and one of the heater elements are on (8) and are using the Red Pitaya to compare it to the data we took for the same measurements on the CyMAC. It will be finished Wednesday morning, and then we plan on starting to take additional recordings in order to conduct the same type of comparison for the dark noise measurements.

Attachment 1: rpon.png
rpon.png
  613   Tue Jul 15 11:47:53 2025 Xuesi MaInfrastructure  

Date and Time: Around 4:50 PM on July 1, 2025

Location and Temperature:

  • Back of the room 1119, around the working station: 85.7 °F
  • Front of the room 1119, around the doorway: 83.5 °F
  • Back of the room 1129, around the working station: 79.1 °F
  • In the hallway in front of room 1119: 76.5 °F
Attachment 1: 20250701_165118.jpg
20250701_165118.jpg
Attachment 2: 20250701_165200.jpg
20250701_165200.jpg
Attachment 3: 20250701_165329.jpg
20250701_165329.jpg
Attachment 4: 20250701_165246.jpg
20250701_165246.jpg
  615   Tue Jul 22 12:20:29 2025 ChristinaUpdateTCS 

[Ma, Christina, Tyler]

Figure 1 shows the following PSDs for channels 0-15 from the 14 hour test. Figure 2 shows the reduced chi squared and CSD plots for both the 2 Hz and 16 Hz resolutions.

Attachment 1: PSDch0-15.png
PSDch0-15.png
Attachment 2: CSD2hz_16hz.png
CSD2hz_16hz.png
  620   Thu Jul 31 14:26:58 2025 Xuesi MaUpdate  

[Tyler, Ma, Christina, Maple, Cece, Mary, Pooyan, Audrey]

We started by cleaning outside of the cleanroom wiping down the cable channel and working our way down while taking the pre-cleaning measurement. We then stated wiping down the inside of clean room and vacuumed, mopped the outside of the cleanroom. Finally, we vacuumed, mopped, and wiped down the floor inside the cleanroom.

Particle Count Measurements:

  • Pre-cleaning (2:00 pm):
    • Zone 3:
      • 0.3 µm: 2707
      • 0.5 µm: 1374
      • 1.0 µm: 249
    • Zone 4:
      • 0.3 µm: 2207
      • 0.5 µm: 1166
      • 1.0 µm: 374
  • Post-cleaning (4:30 pm):
    • Zone 3:
      • 0.3 µm: 4082
      • 0.5 µm: 2415
      • 1.0 µm: 708
    • Zone 4:
      • 0.3 µm: 1707
      • 0.5 µm: 791
      • 1.0 µm: 416
Attachment 1: 2025-08-07_cleanroom_chart.png
2025-08-07_cleanroom_chart.png
  623   Wed Aug 6 16:31:37 2025 Xuesi MaUpdate  

Date and Time: Around 4:20 PM on Auguest 6, 2025

Location and Temperature:

  • Back of the room 1119, around the working station: 90.2 °F
  • Front of the room 1119, around the doorway: 85.8 °F
  • Back of the room 1129, around the working station: 93.4 °F
  • Front of the room 1119, around the doorway: 82.3 °F
  • In the hallway in front of room 1119: 77.2 °F
Attachment 1: 941.jpg
941.jpg
Attachment 2: 942.jpg
942.jpg
Attachment 3: 943.jpg
943.jpg
Attachment 4: 945.jpg
945.jpg
Attachment 5: 946.jpg
946.jpg
  628   Tue Aug 12 13:32:43 2025 LiuUpdateInterferometer Simulations 
Differences in Fused Silica parameters.
Attachment 1: Screenshot_2025-08-12_at_1.31.11_PM.png
Screenshot_2025-08-12_at_1.31.11_PM.png
Attachment 2: Screenshot_2025-08-12_at_1.31.17_PM.png
Screenshot_2025-08-12_at_1.31.17_PM.png
  629   Tue Aug 12 13:46:56 2025 ChristinaUpdate  
Attachment 1: 3.png
3.png
Attachment 2: 3-1.png
3-1.png
  641   Tue Sep 9 12:01:38 2025 Tyler, MaUpdate  
Slides
  644   Tue Sep 9 13:08:44 2025 Xuesi MaUpdate  
Slides
  660   Thu Nov 13 12:39:01 2025 MaryUpdateInterferometer Simulations 
Attachment 1: PA.pdf
PA.pdf
  671   Thu Jan 22 13:12:29 2026 MaryUpdate  
Attachment 1: PA_matem2.pdf
PA_matem2.pdf
Attachment 2: PA_maxtem6.pdf
PA_maxtem6.pdf
Attachment 3: maxtem600.jpg
maxtem600.jpg
Attachment 4: maxtem2dis.jpg
maxtem2dis.jpg
Attachment 5: dismaxtem6.jpg
dismaxtem6.jpg
  681   Thu Feb 5 13:59:42 2026 MaryUpdateInterferometer Simulations 
Attachment 1: 112.pdf
112.pdf
ELOG V3.1.3-7933898