ligo-ex ligo-ds
  Richardson Lab Experimental Log, Page 1 of 13  Not logged in ELOG logo
New entries since:Wed Dec 31 16:00:00 1969
IDdown Date Author Type Category Subject
  630   Tue Aug 12 13:49:24 2025 MaryUpdate point absorber and coated aperture in A#
  629   Tue Aug 12 13:46:56 2025 ChristinaUpdate  
  628   Tue Aug 12 13:32:43 2025 LiuUpdateInterferometer Simulations 
Differences in Fused Silica parameters.
  627   Tue Aug 12 13:28:00 2025 TylerUpdateTCSRH Fenicsx v. HV
  626   Tue Aug 12 12:35:19 2025 TylerUpdate CyMAC ADC Update
[Christina, Ma, Tyler]

Updated CyMAC measurement, comparing 260 hrs of measurement time vs. 405 hrs.

  625   Tue Aug 12 12:16:21 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI optimization for A#
Multi-ring FROSTI optimization for A# with static polishing
  624   Mon Aug 11 22:19:50 2025 ChristinaUpdateTCSRed Pitaya Data Streaming

[Ma, Tyler, Christina]

This past week, we were able to set up the data streaming for continuous acquisition to the PCs, and can conduct the initial tests now. We were able to do this by:

  • Installing the correct OS ecosystems to each Red Pitaya
  • Installing the terminal streaming client
  • Debugging formatting and file saving issues by saving the data in a TDMS file on the host PC.
We have started streaming the ADC data on Monday from both boards for 2 days, which are currently in the clean room. Below are the ADC tests we performed before starting the 2-day run. (Time in ms)

  623   Wed Aug 6 16:31:37 2025 Xuesi MaUpdate  

Date and Time: Around 4:20 PM on Auguest 6, 2025

Location and Temperature:

  • Back of the room 1119, around the working station: 90.2 °F
  • Front of the room 1119, around the doorway: 85.8 °F
  • Back of the room 1129, around the working station: 93.4 °F
  • Front of the room 1119, around the doorway: 82.3 °F
  • In the hallway in front of room 1119: 77.2 °F
  622   Tue Aug 5 12:21:04 2025 LiuUpdateInterferometer SimulationsFROSTI A# optimization
A# multi-ring FROSTI optimization
  621   Tue Aug 5 11:25:17 2025 MaryUpdate point absorber and coated aperture in A#

Simulating how point absorber defects on Advanced LIGO mirrors affect cavity performance by modeling a 4km LIGO-like arm cavity with a point absorber fixed at 5cm off-center on the ITM. In this simulation, displaced the laser beam across a 21×21 grid while calculating the deformation, then measuring the resulting power loss in the cavity (~283W circulating power).

  620   Thu Jul 31 14:26:58 2025 Xuesi MaUpdate  

[Tyler, Ma, Christina, Maple, Cece, Mary, Pooyan, Audrey]

We started by cleaning outside of the cleanroom wiping down the cable channel and working our way down while taking the pre-cleaning measurement. We then stated wiping down the inside of clean room and vacuumed, mopped the outside of the cleanroom. Finally, we vacuumed, mopped, and wiped down the floor inside the cleanroom.

Particle Count Measurements:

  • Pre-cleaning (2:00 pm):
    • Zone 3:
      • 0.3 µm: 2707
      • 0.5 µm: 1374
      • 1.0 µm: 249
    • Zone 4:
      • 0.3 µm: 2207
      • 0.5 µm: 1166
      • 1.0 µm: 374
  • Post-cleaning (4:30 pm):
    • Zone 3:
      • 0.3 µm: 4082
      • 0.5 µm: 2415
      • 1.0 µm: 708
    • Zone 4:
      • 0.3 µm: 1707
      • 0.5 µm: 791
      • 1.0 µm: 416
  619   Tue Jul 29 12:30:00 2025 ChristinaUpdateTCS260 hour ADC update

[Ma, Christina, Tyler]

ADC CyMAC 260 hr, 8192Hz sampling rate channels V3 and Vref6

  617   Tue Jul 29 11:45:19 2025 MapleUpdateTCSMeeting update
  616   Tue Jul 29 11:27:57 2025 LiuUpdateInterferometer SimulationsMulti-ring FROSTI optimization for A#
  615   Tue Jul 22 12:20:29 2025 ChristinaUpdateTCS 

[Ma, Christina, Tyler]

Figure 1 shows the following PSDs for channels 0-15 from the 14 hour test. Figure 2 shows the reduced chi squared and CSD plots for both the 2 Hz and 16 Hz resolutions.

  614   Tue Jul 15 13:13:48 2025 ChristinaUpdateTCSPhotodetector optical post assembly
  613   Tue Jul 15 11:47:53 2025 Xuesi MaInfrastructure  

Date and Time: Around 4:50 PM on July 1, 2025

Location and Temperature:

  • Back of the room 1119, around the working station: 85.7 °F
  • Front of the room 1119, around the doorway: 83.5 °F
  • Back of the room 1129, around the working station: 79.1 °F
  • In the hallway in front of room 1119: 76.5 °F
  611   Tue Jul 15 10:40:23 2025 ChristinaUpdateTCS 

[Christina, Ma, Tyler]

Yesterday, Ma and I started taking data for the PSD/CSD measurements while the photodetectors and one of the heater elements are on (8) and are using the Red Pitaya to compare it to the data we took for the same measurements on the CyMAC. It will be finished Wednesday morning, and then we plan on starting to take additional recordings in order to conduct the same type of comparison for the dark noise measurements.

  610   Mon Jul 14 23:50:22 2025 TylerUpdate Cymac ADC update

I've been looking into the performances of individual channels on the Cymac by computing their individual PSDs and corresponding CSDs that show their noise relation to each other. It appears some channels do have lower noise floors than others, and some combinations of these actually do perform similar to the Red Pitaya (showing below the CSD between CHs 3 and 5), although it doesn't look like it's much of an improvement. The best method as of now still appears to be phase-locking two separate ADCs to reduce the correlated noise floor further.

This can be further discussed at the July 22, 2025 group meeting.

  609   Mon Jul 14 20:45:20 2025 Luke SummaryCleanroomHEPA filter performance metrics

[Luke, Luis]

We’ve been looking into methods for evaluating the performance of the HEPA filters in the clean room. Measuring flow rate appears to be a promising approach, since the filter manual specifies a recommended minimum of 70 ft/min.

At the moment, we don’t have a flow rate measurement device on hand. We're currently looking into two options:

  • A standard vane-style anemometer, which is more affordable but may lack sensitivity near the minimum required flow rate.
  • A hot-wire anemometer, which offers better resolution and accuracy at lower speeds, though it's more expensive.

In the next few days, either Luis or I will contact Terra Universal to see if they have any measurement solutions that would integrate well with our clean room setup, or if they can recommend best practices for verifying HEPA performance.

  608   Thu Jul 10 16:41:18 2025 ChristinaUpdateTCS 

[Christina, Ma, Tyler]

Pomona Box completed and is now on the server rack in 1119. Additionally, the photodetectors were connected to the box and set up in front of the heater elements.

  607   Tue Jul 8 12:15:04 2025 TylerUpdateTCSCyMAC ADC Noise Update
[Tyler, Christina, Ma, Mary]

Below is attached the ADC noise floor of two CyMAC channels vs the Red Pitaya.

The frequency resolution of these RIN spectra are 16 Hz, with N_meas = 19,308,426 for the CyMAC, and N_meas = 263,024.

  606   Tue Jul 8 11:46:55 2025 Maple MatznerUpdateTCSUpdate on SURF progress for Jul 8
  605   Mon Jul 7 17:44:04 2025 TylerUpdateGeneralLab Organization
[Tyler, Pooyan, Ma, Cece, Luke, Luis, Christina, Maple]

After acquiring some new equipment for the labs, a few of us spent time on Thursday and Friday sorting through everything and storing them into their appropriate places. New vacuum parts, power supplies, cables, and much more were included. The rooms are now ready for the lab tours commencing Tuesday afternoon.

  604   Mon Jul 7 16:39:18 2025 ChristinaUpdateTCSPomona Box Drawings

Below are the drawings for the pomona box modifications

  603   Tue Jul 1 13:41:36 2025 Xuesi MaUpdate  
[Ma, Liu]

Slides

  602   Tue Jul 1 12:39:42 2025 TylerUpdateTCSA# FROSTI Profile Optimization Update
[Tyler, Liu]

Slides

  601   Tue Jul 1 11:04:00 2025 ChristinaUpdateTCSFROSTI Reconstruction II

[Ma, Tyler, Christina]

Finished external reconstruction of the FROSTI by installing all the pins to the 2 DB 25 connectors. To ensure everything was operating correctly, we did electrical testing by testing continuity and by checking each of the heater elements' RTD and power resistance values and comparing them to what was tabulated in January. Additionally, we organized the wires and added the stands.

  600   Fri Jun 27 15:38:22 2025 ChristinaUpdateTCSFROSTI Reconstruction

[Tyler, Ma, Christina, Luke]

We threaded the wiring of the heater elements through the reflective surfaces and were able to attach each piece with the Macor spacers and connect both sides. We used the guide rails and screws to ensure proper alignment. The only necessary step left is to add the additional external screws for the reconstruction.

  599   Wed Jun 25 15:44:47 2025 Luke UpdateCleanroomCleanroom cleaning

[Luke, Luis, Tyler, Ma, Christina, Maple]

We started by cleaning outside of the cleanroom wiping down the cable channel and working our way down. We replaced the air filter in the HEPA filter with what seemed to be the last one in storage. We then vacuumed outside, before wiping down the inside of the cleanroom. We then vacuumed, mopped, and wiped down the floor inside the cleanroom.

We are ready for venting the vacuum chamber which is planned to take place 6/26/25 at 10am. 

Particle Count Measurements:

  • Pre-cleaning (12:00 pm):
    • Zone 3:
      • 0.3 µm: 2517
      • 0.5 µm: 662
      • 1.0 µm: 176
    • Zone 4:
      • 0.3 µm: 177
      • 0.5 µm: 44
      • 1.0 µm: 0
  • Post-cleaning (1:45 pm):
    • Zone 3:
      • 0.3 µm: 619
      • 0.5 µm: 88
      • 1.0 µm: 0
    • Zone 4:
      • 0.3 µm: 177
      • 0.5 µm: 88
      • 1.0 µm: 0
  598   Tue Jun 24 13:47:58 2025 LiuUpdateInterferometer SimulationsAlternative FROSTI A# optimization approach
Alternative FROSTI A# optimization approach
  597   Tue Jun 24 12:45:16 2025 TylerUpdateGeneralUpdated ADC spectra
[Tyler]

Updated ADC noise spectra measurements using diaggui.

NOTE: Will update plots with proper axes

  596   Tue Jun 24 11:50:15 2025 Xuesi MaUpdate  
Heater Element Resistance Test Results

Heater Element Resistance Test Report

Test Overview

Conducted several on-and-off tests on heater elements at 24V, followed by switching heater elements 5 and 8 by changing their connection pins to the ADC.

Resistance Measurements (Ω)

Heater Element Before Switch After Switch Ladder Test Difference (After-Before)
Element 1 153.09577918 153.11447008 153.83373894 +0.0186909
Element 2 149.70967557 149.8175605 150.08915205 +0.10788493
Element 3 157.88314236 157.87136021 157.39696467 -0.01178215
Element 4 146.80153622 146.78826747 147.77923177 -0.01326875
Element 5 151.80888657 152.2204132 150.44919044 +0.41152663
Element 6 148.24375881 148.32846152 146.89161192 +0.08470271
Element 7 144.72286546 144.76879963 145.03627875 +0.04593417
Element 8 151.4564267 149.14240417 149.88675562 -2.31402253

Supporting Graphs

Attached below are the resistance-time graphs for all heater elements:

  • Complete test duration graph
  • Pre-switch period graph
  • Post-switch period graph
  • Resistance voltage graph after removing data from after power outage

Analysis & Conclusion

The switch did not physically change the resistance of the heater elements, suggesting the measurements reflect their true resistances.

  595   Tue Jun 17 12:46:47 2025 TylerUpdateElectronicsCyMAC ADC Noise
Below are results from power spectra measurements of the CyMAC ADC, computed form 30 hours of time series data, compared to the performance of the Red Pitaya. The frequency resolution for the initial Cymac CSD is 2 Hz, and 216,000 spectra measurements are averaged together. As of now, it still looks like we haven't reached the correlated noise floor of the CyMAC. We have a few days worth of data to take from, so I'll look into using a longer period of time.
  594   Tue Jun 17 12:22:43 2025 Xuesi MaUpdate  
CYMAC Remote Control Cable Documentation

CYMAC Remote Control Cable Assembly

Cable Functions

The custom cable assembly serves two primary functions:

1. Jumper Connections (DB25 Side - Connector J3)

The following pins are jumpered on the DB25 connector:

From Pin To Pin
Pin 3 Pin 16
Pin 10 Pin 11
Pin 12 Pin 24
Pin 13 Pin 25
Pin 22 Pin 23

2. Signal Wiring (DB9 to DB25)

The following connections are made between the DB9 and DB25 connectors:

DB9 Pin DB25 Pin Function
Pin 1 Pin 9 Positive side
Pin 6 Pin 12 Negative side
Pin 5 Pin 6 Ground

Note: All connections are verified with a multimeter.

Note: This cable allows remote control of the Sorensen power supply with an external voltage source (Cymac DAC).

Note: For more details, check section 2.11.1 from the attached Sorensen power supply manual.

  593   Wed Jun 11 16:33:45 2025 PooyanUpdateComputersLuke's ray tracing hosting
  • Hosted Lukes model on Chimay. It's available at https://richardsonlab.ucr.edu/real-time-frosti
  • The sourcecode is at Github. It is clones at /var/www/html/real-time-frosti, and this is the block added to nginx config file at /etc/nginx/sites-available/default :

    location /real-time-frosti {

    root /var/www/html;
    index index.html index.htm index.php;
    }
  • Also added this line to the crontab, so the code will be checked for updated from the sourcecode every five minutes. 

    */5 * * * * cd /var/www/html/real-time-frosti/ && git pull 

  • TODO: Move the sourcecode from Github to git.ligo.org, and make the repository public. 
  592   Wed Jun 11 16:19:37 2025 PooyanUpdateCDSCDS update
  • Fast channels recorded to .gwf frame files
    • Had to add a DAQ Channels block to the c1msc model for those channel to be recorded in the .gwf frame files. The block is in the CDS_PARTS, and I just had to copy it to our model. Here is the example text for the channels:

      #DAQ Channels

      ONE_DAQ_CHANNEL 2048
      ANOTHER_DAQ_CHANNEL 1024
      SCIENCE_FRAME_CHAN* 1024
      UINT32_CHAN uint32 2048
      DAQ_CHANNEL_AT_DEFAULT_RATE

      I added two channels to the list (V0 and V1). Remember that the channel name should not be the full name, but just the part's name. In this case, I added these two channel names:

      V0_OUT
      V1_OUT

      with the default sample rate. 
  • The fast channels are recorded with an extra _DQ in their names. So, in ndscope and diaggui these two channels should be addressed as
    • C1:MSC-V0_OUT_DQ
    • C1:MSC-V1_OUT_DQ
  • Update 6/13: enabled recording of all the ADC voltage channels so that Tyler can use all the data for the noise floor calculation. attached is an example of diagggui with almost 30 hrs of data.
  • TODO: Enable daqd_wiper
  591   Tue Jun 10 12:13:00 2025 Xuesi MaUpdate  

By measuring the resistance of the ladder, I obtained a graph of resistance as a function of voltage.
All the heater elements behave consistently, showing only a small standard deviation.
However, heater element 5 shows a significantly larger standard deviation.
Upon examining the initial ladder graph, I noticed that the resistance of heater element 5 increases following a power outage that resets the system.
At this point, I am unsure why this behavior occurs.

  590   Tue Jun 10 12:11:14 2025 PooyanUpdateCDSCDS update
  • The MEDM models issue was fixed. The issue was that building the c1msc model alone was not enough and we had to re-build the c1iop as well, although no changes has been made to the c1iop. Added a note in the wiki on this for future reference.
  • For writing the fast channels to the frame files, our CDS system should have a GDS broadcaster. I followed this wiki page to set one, but it breaks the current daqd system. Asked about it earlier today on the CDS Mattermost channel and waiting for a response. 
  • Maybe I should continue the distributed scheme we have been trying for the nds2-server and let another machine on the network do this. 
  • Question: daqd has a feature to zero-out the bad data (NaNs), and it is currently on for our system. I don't think that it's a good idea to have this. Should I turn it off? 
  589   Tue Jun 3 13:16:53 2025 Xuesi MaInfrastructure  

Date and Time: Around 3:50 PM on June 2, 2025

Location and Temperature:

  • Back of the room 1119, around the working station: 86 °F
  • Front of the room 1119, around the doorway: 84.1 °F
  • Back of the room 1129, around the working station: 78.6 °F
  • Front of the room 1129, around the doorway: 78.3 °F
  • In the hallway in front of room 1119: 75.2 °F
  588   Tue Jun 3 13:08:47 2025 Xuesi MaUpdate  

We currently have the parts for a DB9 connector but need parts for a DB25 connector, where did we get the parts for the DB9 connector? We may be able to use the same company for the DB25 connector.

Also need a cable for the connectors. Find a DB9 to DB9 cable in 1129, not sure if I can disassemble it and use just the cable.

The pressure in the vacuum chamber rises to 2.12e-8 torr, which needs baking.

  587   Tue Jun 3 12:16:13 2025 Cynthia UpdateInterferometer SimulationsPoint absorber simulation on radially averaged profile
  586   Tue Jun 3 11:54:06 2025 Cece OchoaUpdate CE Cavity Design Update
Update on the design for the CE Cavity Mirrors
  585   Tue Jun 3 11:50:08 2025 PooyanUpdateCDSCDS update

There are three major issues with the lab's CDS system that should be addressed. They are ordered from the highest priority to the lowest.

  • CyMAC has stopped serving the fast channels (65k Hz)
    The channels are still there, and the data from the slow (16 Hz) channels is served correctly, but the fast data is not served. As a result, diagggui can not retrieve any information at the moment.
    Probably the issue is the same as the version conflict between daqd and nds2-server. After installing nds2-server on CyMAC and realizing that the conflict can not be resolved, I reverted the changes so that the CyMAC could be compatible with daqd, but there is a chance that there are still version conflicts that block the fast channels.
    I suspect that we have broken the MSC model when adding the new button. This could explain both the fasr-channels fault and the unexpected crashes.

    UPDATE 6/5
    Fixes this by rebuilding the models.
    Note for the future: The c1iop and c1msc models should be rebuilt together, even if no change has been made to c1iop.
  • The fast channels were not saved to the frames
    Reviewing the frame files showed that the daqd was only saving the 16 Hz data in the .gwf frame files. I had missed this because I only checked the existence and the integrity of the raw frame files, but never retrieved the high-frequency data itself to check it.
  • NDS2-server is installed, but does not distribute the data
    I managed to install the nds2-server and its satellite programs on the Chimay, and NFS mount the frame files on it. The server successfully chaches the channel names, frame files, and the data, but fails to list the channels for distribution.

Note: CyMAC machine was strangely disconnected from the network, and required manual reboot. I couldn't find any abnormal logs that could explain what happened.

  584   Mon Jun 2 09:45:06 2025 Luke UpdateScripts/ProgramsRingheater modeling Update

Slides

I have added some slides to my meeting updates. The updates include: my choice in position and width, some figures I have added to my write up, and what I think my immediate steps are to wrap up.

I have also made a new FROSTI model that conducts its raytracing in real time. Something that was severely lacking in the previous model.

  583   Tue May 27 12:52:22 2025 Xuesi MaUpdate  

Using the configuration in section 2.11.1 in the Sorensen installation and operation manual (attached below), I'm able to remote control the Sorensen power supply.

I use DAC VEXC8 as the external voltage source and a breakout board to match the configuration of connector J3.

I have recorded 7 cycles (attached below), and am now working on analyzing the data.

  582   Mon May 26 23:24:33 2025 MichaelUpdateInterferometer SimulationsInitial Monte Carlo Simulations for Toroidal Mirrors Project
I've run an initial Monte Carlo simulation for statistical comparison of toroidal mirrors for the dual FROSTI TCS case. Below is the link to the slides.

slideshow

  581   Mon May 26 11:15:28 2025 LukeUpdateScripts/ProgramsUpdate to optimization plots
I have created some new optimization plots. I have also started on a model like the previous Desmos model, but in python.

Slides

  580   Wed May 21 14:40:38 2025 Xuesi MaInfrastructureCleanroomCleanroom Temperature

Date and Time: Around 2:20 PM on May 21, 2025

Location and Temperature:

  • Back of the room, around the working station: 85.5 °F
  • Front of the room, around the doorway: 82.3 °F
  579   Mon May 19 10:59:14 2025 Xuesi MaInfrastructure  

Status Update

Previous Status:

All Sorensen power supplies were turned off. The Cymac system was down (all chassis were offline).

If you need more information or assistance in turning them back on, please contact Xuesi Ma.


Update: The systems are now back online.

Note: The Sorensen power supplies’ voltage can now be controlled via channel VEXC8 of the DAC.

If scripting is required, the ezca Python package is highly recommended for interfacing and control.

ELOG V3.1.3-7933898