ID Date Author Type Categorydown Subject
  553   Tue Apr 15 11:13:43 2025 Xuesi MaUpdate  

Test 4: AI Chassis ground isolation

  • Setup: AI Chassis isolated from the ground with readout directly connected to adapter board
  • Observation: No spike observed.

Test 5: Ribbon cable check

  • Setup: AI Chassis remains isolated from the ground. The signal is measured after the ribbon cable. (Attachment 3)
  • Observation: No spike observed.

Test 6: Single ground connection check

  • Setup: AI chassis powered on but remains isolated from the frame. The signal is measured from the output of the AI Chassis. (Attachment 1 ,2)
  • Observation: No spike observed.
Attachment 1: 20250414_120428.jpg
20250414_120428.jpg
Attachment 2: 20250414_120423.jpg
20250414_120423.jpg
Attachment 3: 20250411_163105.jpg
20250411_163105.jpg
  559   Tue Apr 22 12:08:46 2025 Xuesi MaUpdate  

Test 7: Reconnect ribbon cables

  • Setup: Reconnect all the ribbon cables, and turned on the AI Chassis while isolated rack
  • Observation: No spike observed.

Test 8: Reconnect to rack

  • Setup: Reconnect the AI Chassis to the rack.
  • Observation: No spike observed.

Test 9: Restore to original position.

  • Setup: CLose up the AI Chassis, and put it back to its original location.
  • Observation: No spike observed.
Attachment 1: 20250415_161755.jpg
20250415_161755.jpg
Attachment 2: 20250416_174640.jpg
20250416_174640.jpg
Attachment 3: 20250422_105614.jpg
20250422_105614.jpg