ligo-ex ligo-ds
  Richardson Lab Experimental Log  Not logged in ELOG logo
Message ID: 529     Entry time: Tue Mar 4 03:33:18 2025
Author: Xuesi Ma 
Type: Update 
Category:  
Subject: Update 

Spikes Appear Again, need to address it systematically.

Power on and off before reaching steady state ✔

At 12V, the rise and fall time of heater elements are different from 24V.

Initial guess is due to temperature in the chamber. However, it does not seem to be the case, 24V with 30c have the same hall time as 24V at lower temperature.

Attached are some graphs for rise and fall time

After a weekend of powering on, the main chamber pressure stabilized in the UHV region.

Temperature in the chamber seems also not to change.

Attachment 1: resistance_vs_time_all_channels_all_time.pdf  677 kB  | Hide | Hide all | Show all
Cannot create thumbnail, please check ImageMagick installation
Attachment 2: resistance_vs_time_rise_time.pdf  154 kB  | Hide | Hide all | Show all
Cannot create thumbnail, please check ImageMagick installation
Attachment 3: resistance_vs_time_fall_time.pdf  279 kB  | Show | Hide all | Show all
Attachment 4: resistance_vs_time_rise_time_12V.pdf  270 kB  | Hide | Hide all | Show all
Cannot create thumbnail, please check ImageMagick installation
Attachment 5: resistance_vs_time_fall_time_12V.pdf  174 kB  | Hide | Hide all | Show all
Cannot create thumbnail, please check ImageMagick installation
Attachment 6: resistance_vs_time_fall_time_24V_30c.pdf  215 kB  | Hide | Hide all | Show all
Cannot create thumbnail, please check ImageMagick installation