This corner plot was generated using the base finesse-ligo LLO model for the MCMC forward model. The prior distributions were centered around the base model's parameter values. Then each parameter was slightly perturbed according to their prior width and this perturbed instance of the model was used to generate a list of detector values. Each detector reading was then perturbed with 1% fractional error to simulate detector/calibration uncertainties. From the plot, the sampler has a hard time recovering the test mass RoC's however it can recover the PRC mirror positions well.
Entry  Mon May 1 17:10:25 2023, Julian, Cao, Physics, VAC, Installing Pirani gauge and RGA probe onto to vacuum chamber PXL_20230501_225241491.jpgPXL_20230501_235854958.jpgPXL_20230501_234750795.MP.jpg

[Julian, Cao]

  • 03:00 pm: Particle count
    1. : Zone 3
      • 0.3 um: 498
      • 0.5 um: 124
      • 1 um: 124
    2. : Zone 4
      • 0.3 um: 748
      • 0.5 um: 457
      • 1 um: 374
  • 03:32 pm: Installing Pirani gauge onto pump line
  • 03:51 pm: Finish installing Pirani gauge, start installing RGA probe
  • 04:03 pm: Finish installing PRGA probe, start routing cable from gauge controller to Pirani gauge
  • 04:30 pm: All ethernet cables are routed and connected to gauge controllers
  • 04:48 pm: Power gauge controller up, all gauges are recognised and readout shows atmospheric pressure as expected (1000 mbar)
  • 04:51 pm: End-of-work particle count
    1. : Zone 3
      • 0.3 um: 1413
      • 0.5 um: 872
      • 1 um: 623
    2. : Zone 4
      • 0.3 um: 374
      • 0.5 um: 166
      • 1 um: 166
We also closed all the valves so we can start testing vacuum pump down of isolated volume tomorrow.