ligo-ex
ligo-ds
|
| Richardson Lab Experimental Log |
Not logged in |
 |
|
|
Message ID: 767
Entry time: Mon Nov 16 12:43:44 2026
|
| Author: |
Shane |
| Type: |
Update |
| Category: |
Cleanroom |
| Subject: |
Cleanroom Cleaning + Safety Concerns |
|
|
[Shane, Cece, Ma, Mary, Ruaraidh]
The cleanroom has now been fully cleaned, and floors have been vacuumed in all three lab rooms, and mopped in 1119 and 1125. Aside from cleaning, we also noticed a number of potential issues:
- Spotted a number of ants in 1119 again
- All three labs were especially hot today, despite it being colder than usual outside. We used the anemometer to measure the temperature of the air coming out of the main vents in the ceiling, and air is being pushed into the room at 86.2 degrees F (1119), 87.4 degrees F (1125), and 88.9 degrees F (1129). Images of readings attached.
- There is a screw bending under the optical table in the cleanroom on the side where the vacuum chamber is. Image attached.
- There's new oily looking marks on the yellow vacuum chamber cover where the heating tape is. No residue lifted when wiped and the texture was a little rougher than the rest of the vacuum cover. There's also a section of the heating tape that seems to be melting into the foam cover by the connection point at the bottom of the RGA. Images attached in PDF.
|
|
|
|
|
|
|
|
|
|
|