ligo-ex
ligo-ds
|
| Richardson Lab Experimental Log |
Not logged in |
 |
|
|
Message ID: 763
Entry time: Fri Nov 6 10:17:04 2026
|
| Author: |
Shane |
| Type: |
Infrastructure |
| Category: |
Cleanroom |
| Subject: |
Ceiling Fan Flow Rate Tests |
|
|
[Shane, Ma, Mary, Cece]
We've finished 6-pt flow rate tests for all four cleanroom ceiling fans. Results are attached as pdf. All of the 6-pt tests were done with the fan speed on high, as the anemometer is only rated to resolve values above 80 fpm. Also attached here as a photo is a clearer digram of the measurement scheme, with description in the PDF.
|
|
|
Cannot create thumbnail, please check ImageMagick installation
|
|
|