Message ID: 634     Entry time: Tue Aug 26 12:18:36 2025
Author: Michael 
Type: Update 
Category: Interferometer Simulations 
Subject: Squeezing Plots for SRC Parameter Space 
I've run a few plots of observed squeezing in the LIGO interferometer to fully ensure that the point we have selected is the optimal point for squeezing. Contrary to what I expected, there seems to be much more degenerate "optimal" configurations. While the toroidal case may be near optimal, it seems that the solution we found for a non astigmatic beam is not the only one. I will further explore this space to find more optimal/tolerant positions, which I see evidence of. Additionally, as observed squeezing is computational expensive to compute, I found that the difference in round-trip gouy phase accumulation between the x and y planes is a good approximation for an optimal squeezing state, which I believe makes physical sense. This may be helpful to quickly compute other optimal conditions.
Attachment 1: Screenshot_2025-08-26_at_12-16-01_LIGO_Project_Log_-_Google_Docs.png  162 kB  | Show | Hide all | Show all
Attachment 2: Screenshot_2025-08-26_at_12-16-32_LIGO_Project_Log_-_Google_Docs.png  182 kB  | Hide | Hide all | Show all
Screenshot_2025-08-26_at_12-16-32_LIGO_Project_Log_-_Google_Docs.png
Attachment 3: Screenshot_2025-08-26_at_12-17-26_LIGO_Project_Log_-_Google_Docs.png  186 kB  | Hide | Hide all | Show all
Screenshot_2025-08-26_at_12-17-26_LIGO_Project_Log_-_Google_Docs.png