ligo-ex ligo-ds
  Richardson Lab Experimental Log  Not logged in ELOG logo
Message ID: 596     Entry time: Tue Jun 24 11:50:15 2025
Author: Xuesi Ma 
Type: Update 
Category:  
Subject:  
Heater Element Resistance Test Results

Heater Element Resistance Test Report

Test Overview

Conducted several on-and-off tests on heater elements at 24V, followed by switching heater elements 5 and 8 by changing their connection pins to the ADC.

Resistance Measurements (Ω)

Heater Element Before Switch After Switch Ladder Test Difference (After-Before)
Element 1 153.09577918 153.11447008 153.83373894 +0.0186909
Element 2 149.70967557 149.8175605 150.08915205 +0.10788493
Element 3 157.88314236 157.87136021 157.39696467 -0.01178215
Element 4 146.80153622 146.78826747 147.77923177 -0.01326875
Element 5 151.80888657 152.2204132 150.44919044 +0.41152663
Element 6 148.24375881 148.32846152 146.89161192 +0.08470271
Element 7 144.72286546 144.76879963 145.03627875 +0.04593417
Element 8 151.4564267 149.14240417 149.88675562 -2.31402253

Supporting Graphs

Attached below are the resistance-time graphs for all heater elements:

  • Complete test duration graph
  • Pre-switch period graph
  • Post-switch period graph
  • Resistance voltage graph after removing data from after power outage

Analysis & Conclusion

The switch did not physically change the resistance of the heater elements, suggesting the measurements reflect their true resistances.

Attachment 1: resistance_vs_time_all_channels_channel_switch.jpg  63 kB  | Hide | Hide all | Show all
resistance_vs_time_all_channels_channel_switch.jpg
Attachment 2: resistance_vs_time_all_channels_channel_switch_one_cycle_before.jpg  58 kB  | Show | Hide all | Show all
Attachment 3: resistance_vs_time_all_channels_channel_switch_one_cycle_after.jpg  58 kB  | Hide | Hide all | Show all
resistance_vs_time_all_channels_channel_switch_one_cycle_after.jpg
Attachment 4: resistance_vs_voltage.jpg  45 kB  | Hide | Hide all | Show all
resistance_vs_voltage.jpg
ELOG V3.1.3-7933898