Message ID: 574     Entry time: Fri May 9 20:06:27 2025
Author: Luke  
Type: Summary 
Category: Cleanroom 
Subject: Cleanroom cleaning 

[Luke, Mohak, Luis]

We were also able to remove the residue that has built up on the soft walls by donning a skinny bunny suit and with two wipes rubbing both sides of the soft walls at the same time. The wipes would get gummy very quickly and need to be replaced frequently.

Particle Count Measurements:

  • Pre-cleaning (1:20 pm):
    • Zone 3:
      • 0.3 µm: 3577
      • 0.5 µm: 1369
      • 1.0 µm: 485
    • Zone 4:
      • 0.3 µm: 1062
      • 0.5 µm: 442
      • 1.0 µm: 177
  • Started Cleaning (3:00 pm)
  • Finished Cleaning (4:00 pm)
  • Post-cleaning (4:22 pm):
    • Zone 3:
      • 0.3 µm: 3577
      • 0.5 µm: 618
      • 1.0 µm: 132
    • Zone 4:
      • 0.3 µm: 2031
      • 0.5 µm: 397
      • 1.0 µm: 132

 

Attachment 1: 20250509_particalcount.png  40 kB  | Hide | Hide all