Message ID: 566     Entry time: Wed Apr 30 15:43:56 2025
Author: Luis 
Type: Update 
Category: VAC 
Subject: Vacuum Update 

[Ma, Luke, Luis]

April 29, 2025

3:30 PM
We performed another leak test on the vacuum system, this time focusing on the welds around the main cylinder. The leaks from these junctions were in the mid 10⁻¹¹ range, but we noticed some strange behavior after completing the test. The partial pressure of He kept rising at a slow, constant rate for a while—even after stopping the test for about 15–20 minutes and restarting the measurement. We're not sure what’s causing this and wanted to check in before moving forward with the bake.

4:30 PM
We performed an RGA of the system (see below) and the findings are that there was a higher contraction of He than usual. We have attached another scan from 2/24/25 for reference.

5:00 PM
We turned off the electron multiplier and closed the RGA Program.

Attachment 1: 2025-4-29_HeaterElement-Bake1_ArO.png  328 kB  | Hide | Hide all
2025-4-29_HeaterElement-Bake1_ArO.png
Attachment 2: 2025-2-24_HeaterElement_ArO.png  316 kB  | Hide | Hide all
2025-2-24_HeaterElement_ArO.png