Message ID: 557     Entry time: Tue Apr 22 08:00:39 2025
Author: Luke  
Type: Update 
Category: Scripts/Programs 
Subject: Thermal optimization modeling 40kg 

I have redone the width and location optimization with a 40kg test mass. This was to see where the nominal ring heater would fall in this optimization map see attached.

There are two concerns I have with this modeling:

  1. I am using normalized power only imparting 1W, which may effect the single pass scattering used for optimization.
  2. I am only looking at the surface deformation and I believe that looking at OPD may also be something that is important to consider. 

Note z_RH is measured from the HR surface and w_RH is the half width of the target plane