Upon final review of the FROSTI analysis included in the (hopefully) soon-to-be submitted instrument paper, I've made some adjustments to the reflectivity analysis that estimates the amount of power delivered to the test mass by the FROSTI. Initially, as detailed in elog 447, the delivered power was approximated to be 12.6 W (later adjusted to 12.9 W), with roughly 18% of this power being reflected by the test mass. After some final adjustments, this value now is closer to 12.0 W, with 10.2 W absorbed (15% power reflection). Below is a table showing the updated values for the FROSTI prototype tests:
|
Old Value |
New Value |
Delivered Power |
12.9 W |
12.0 W |
Absorbed Power |
10.6 W (18% reflectivity) |
10.2 W (15% reflectivity) |
Measured Peak Temperature Difference |
5.39 +/- 0.03 K |
5.26 +/- 0.03 K |
Model Peak Temperature Difference |
5.42 K |
5.21 K |
Measured Peak OPD |
765 +/- 9 nm |
771 +/- 7 nm |
Model Peak OPD |
630 nm |
654 nm |
|