|
ligo-ex
ligo-ds
|
| Richardson Lab Experimental Log |
Not logged in |
 |
|
|
Message ID: 54
Entry time: Wed Mar 22 15:25:00 2023
|
| Author: |
Shane |
| Type: |
Update |
| Category: |
General |
| Subject: |
Clean room recertification particle counts |
|
|
| Attached are the five zone particle measurements for the clean room now that the fan speeds have been set to MEDIUM. 2 min sample time, 3 samples per zone. The occupied clean room measurements were unusual in that they generally got larger towards minute 6 instead of settling and going to zero, and for this reason didn't stay beneath the class 5 requirement at the end of the samples. The unoccupied counts look more similar to what was expected, and generally settle under the class five requirement. |
|
|
Cannot create thumbnail, please check ImageMagick installation
|