ligo-ex
ligo-ds
|
Richardson Lab Experimental Log |
Not logged in |
 |
|
Message ID: 516
Entry time: Tue Feb 18 10:12:39 2025
|
Author: |
Tyler |
Type: |
Update |
Category: |
TCS |
Subject: |
Final RIN Plot Update |
|
|
Below is a preview of the final RIN figure that will be included in the FROSTI instrument paper. A quick summary of what is shown below:
The original RIN CSD measurement is shown on the top panel in red. Frequency bins that exhibit external electronics noise (ex. ADC, photodetector noise, etc.) are identified and shaded in gray. These noisy bins are then excluded from the dataset before beginning the next step in the analysis process: rebinning. Here, the resolution of the spectrum can be changed by averaging frequency bins together within a specified interval, with the goal of pushing the measurement curve closer to the A+ requirement shown in the figure. For demonstration below, the spectrum goes from a resolution of ~0.93 Hz to 14.90 Hz.
|
|
Cannot create thumbnail, please check ImageMagick installation
|
|
Cannot create thumbnail, please check ImageMagick installation
|
|
Cannot create thumbnail, please check ImageMagick installation
|
|
Cannot create thumbnail, please check ImageMagick installation
|