[Luke, Luis]
After the gasket replacement performed on Monday (11/18/24) we let the vacuum pump down to UHV pressure for a couple days. Today (11/20/24) we measured the pressure to be 5.30*e^-9 with a temperature of 22C, and we performed a leak test and RGA Scan.
Here are the results:
Connection |
Leak rate (torr) |
Reducing Cross - TP Left |
4.5e-9 (lowish) |
Reducing Cross - TP Right |
6.0e-9 (lowish) |
|