ligo-ex ligo-ds
  Richardson Lab Experimental Log  Not logged in ELOG logo
Message ID: 474     Entry time: Tue Nov 19 17:10:34 2024
Author: Luke  
Type: Update 
Category: VAC 
Subject: Leak check 

Chamber temperature: 23C

Chamber Initial Pressure: ~1.8e-8 main, ~3.5e-8 rga

Did another leak test after monday's work to check how the seal was holding up initial leak was 5.5e-9. I then tried to tighten the bolts a bit more the leak rose to 1.9e-8 about a factor of 4 change.

I checked another flange to confirm that I was preforming the test properly. I measured the elbow to pressure sensor to be 2.2e-11 aproximatly what it was earlier in the month (1.9e-11). 

I ran another isolation test. It is still exponetial. Graph attached.

Thoughts:

I belive at this point that it is a knife edge issue when replacing the gasket I checked both flanges and the new gasket. The tp knife edge seemed perfectly fine, the gasket had very minor scratches, and the reducing cross had very small imperfections on its edge. 

We may want to try and use the liquid sealant or get annealed gaskets for 4.5"

Attachment 1: image.png  38 kB  | Hide | Hide all | Show all
image.png
Attachment 2: IMG_5116.jpeg  2.256 MB  Uploaded Wed Nov 20 12:07:40 2024  | Show | Hide all | Show all
Attachment 3: IMG_5117.jpeg  1.974 MB  Uploaded Wed Nov 20 12:07:57 2024  | Show | Hide all | Show all
ELOG V3.1.3-7933898