  Richardson Lab Experimental Log  Not logged in ELOG logo
Message ID: 386     Entry time: Wed Jun 26 16:39:34 2024
Author: Luke 
Type: Update 
Category: VAC 
Subject: Vacuum chamber leak checking  
On Tuesday I had isolated the main volume and let it pump down over-night. Today, Wednesday, I checked it and it had only gone down by a small amount compared to when I left it on Tuesday. I then proceeded to tighten down the bolts attaching the tubo-pump to the zero-length reducer while it was under vacuum. This caused the pressure to drop down a few orders of magnitude.
Attachment 1: image_2024-06-26_162539705.png  2.660 MB  | Hide | Hide all | Show all
image_2024-06-26_162539705.png
Attachment 2: image_2024-06-26_162601616.png  2.342 MB  | Show | Hide all | Show all