ligo-ex ligo-ds
  Richardson Lab Experimental Log  Not logged in ELOG logo
Message ID: 306     Entry time: Tue Jan 16 16:11:41 2024
Author: Aiden 
Type: Update 
Category: VAC 
Subject:  
[Aiden,Luis]

Below is the data for the most recent bake. I believe the chamber is cleaner than it is actually showing as the RGA data was still getting lower while the data was being taken. I will take the data again tomorrow just in case. Still, the data shows that the chamber is basically as clean as it was before bake 10 and this means further power testing can proceed.

Helium leak tested the chamber again today to check if the sealant changed anything. The two primary target flanges from last time have improved significantly with the flange from the RGA on the cross now being below 2e-14. While the turbo pump flange was leaking at a rate of 4e-13. Much better than what it has been before.

Attachment 1: bake10_240116_Elements_ArO.png  298 kB  | Hide | Hide all | Show all
bake10_240116_Elements_ArO.png
Attachment 2: bake10_240116_Elements_ArC.png  306 kB  | Show | Hide all | Show all
ELOG V3.1.3-7933898