ligo-ex ligo-ds
  Richardson Lab Experimental Log  Not logged in ELOG logo
Message ID: 264     Entry time: Mon Nov 13 11:07:50 2023
Author: Tyler 
Type: Update 
Category: VAC 
Subject: RTD Analysis 

After initial analysis from last week on a single RTD, I then extended to looking at all 8 in series with R_ref (set to 1 kOhm). Shown below are the edge cases for the setup:

  • RTDs are all at ambient lab temperature. This would correspond to a minimum resistance value.
  • RTDs all read out 400 C. This gives the maximum resistance value.

The results show that indeed only a few mA of current is drawn even at room temperature (a little above 5.5 mA), and this will continue to decrease with increasing temperature. The voltage across a single terminal, at a maximum, is only about 5.4 V.

Attachment 1: Screenshot_2023-11-13_at_11.05.45_AM.png  66 kB  | Show | Hide all | Show all
Attachment 2: Screenshot_2023-11-13_at_11.06.04_AM.png  70 kB  | Hide | Hide all | Show all
Screenshot_2023-11-13_at_11.06.04_AM.png
ELOG V3.1.3-7933898