ligo-ex ligo-ds
  Richardson Lab Experimental Log  Not logged in ELOG logo
Message ID: 213     Entry time: Thu Aug 17 15:18:48 2023
Author: Aiden 
Type: Update 
Category: VAC 
Subject: VAC Heating System Update 
Relocated the PID controllers so that the heating tape could be plugged in. In the picture the left one controls the top half of the chamber along with the RGA line and has its sensor on the lid near a fold. The right one controls the bottom half along with the turbo pump line and the bypass line and has its sensor on the barrel of the chamber. The chamber is almost ready to bake again and just needs to have the full range gauges and RGA electronics box removed.
Attachment 1: IMG_4628.jpg  139 kB  | Hide | Hide all
IMG_4628.jpg
ELOG V3.1.3-7933898