Subject: VAC Heating System Update 

Put the heating tape back on and made sure to have 4 coils around the barrel of the chamber. To make this happen I did not put any heater tape on the ports but made sure to place some of the heater no more than 1 inch away from the ports. Also re-placed some of the heating sensors. For the PID controllers there is one sensor placed on the lid and one sensor placed on the barrel. For the temperature alarms one sensor was placed near the flange of the turbo pump and another placed on the cross in the RGA line. Note that all four of these sensors were placed under the heating tape compared to the last few bakes where they were on top.

The insulation was also put back on and now the chamber should be ready for a low temperature bake after some RGA data is taken before for comparison.

Attachment 1: IMG_4615.jpg  151 kB  | Show | Hide all | Show all
Attachment 2: IMG_4616.jpg  153 kB  | Hide | Hide all | Show all
IMG_4616.jpg
Attachment 3: IMG_4617.jpg  142 kB  | Hide | Hide all | Show all
IMG_4617.jpg
Attachment 4: IMG_4618.jpg  124 kB  | Hide | Hide all | Show all
IMG_4618.jpg
Attachment 5: IMG_4619.jpg  148 kB  | Hide | Hide all | Show all
IMG_4619.jpg
Attachment 6: IMG_4621.jpg  143 kB  | Show | Hide all | Show all
Attachment 7: IMG_4622.jpg  112 kB  | Show | Hide all | Show all
Attachment 8: IMG_4623.jpg  148 kB  | Show | Hide all | Show all
Attachment 9: IMG_4624.jpg  121 kB  | Hide | Hide all | Show all
IMG_4624.jpg
Attachment 10: IMG_4625.jpg  120 kB  | Hide | Hide all | Show all
IMG_4625.jpg
Attachment 11: IMG_4627.jpg  144 kB  | Hide | Hide all | Show all
IMG_4627.jpg
ELOG V3.1.3-7933898