ligo-ex ligo-ds
  Richardson Lab Experimental Log  Not logged in ELOG logo
Message ID: 195     Entry time: Tue Aug 8 15:31:26 2023
Author: Aiden 
Type: Update 
Category: VAC 
Subject: VAC Heating System Update 

Vacuum Chamber equilibrium temperature for bake 3 are 126 C and 88 C for the chamber wall and turbo pump flange respectively. There were reached with the PID controllers both being set to 150 C.

Also noticed some discoloration around the port insulation. It could be burned from being exposed to the heater tape directly without insulation or it could be that the insulation, which looks much different, is depositing something around the ports.

The turbo pump overheated and the heating tape had to be turned off in order to get the pump back online. It was unusually hot at 52 C. Hottest it got in bake 2 was 38 C.

Attachment 1: IMG_4583.jpg  93 kB  | Hide | Hide all | Show all
IMG_4583.jpg
Attachment 2: IMG_4581.jpg  100 kB  | Show | Hide all | Show all
Attachment 3: IMG_4584.jpg  95 kB  | Show | Hide all | Show all
Attachment 4: IMG_4582.jpg  91 kB  | Show | Hide all | Show all
Attachment 5: IMG_4579.jpg  93 kB  | Show | Hide all | Show all
Attachment 6: IMG_4578.jpg  95 kB  | Hide | Hide all | Show all
IMG_4578.jpg
Attachment 7: IMG_4577.jpg  104 kB  | Show | Hide all | Show all
Attachment 8: IMG_4576.jpg  95 kB  | Hide | Hide all | Show all
IMG_4576.jpg
Attachment 9: IMG_4572.jpg  94 kB  | Hide | Hide all | Show all
IMG_4572.jpg
Attachment 10: IMG_4575.jpg  105 kB  | Show | Hide all | Show all
ELOG V3.1.3-7933898