Message ID: 184     Entry time: Tue Aug 1 12:10:03 2023
Author: Aiden 
Type: Update 
Category: VAC 
Subject: Vacuum Chamber Baking Update 

Dr. Richardson and I took off the insulation on the lid again to add more insulation in-between the heater tape folds in addition to what was added on top of the tape after discovering discoloration. Also cleaned the lid with IPA wipes.

Also took more RGA data, one with the leak closed and one with the leak open. The graphs below this show that the SNR is improving with the ratio now looking to be around 3. The chamber is still too dirty to be commissioned and we will need to be baked again soon.

Plugged the temperature sensor back in after getting another extension cord. Removed the electronics box for the RGA as well as the full range gauge above the RGA line.

Attachment 1: bake2_230801_ArC.png  334 kB  | Show | Hide all | Show all
Attachment 2: bake2_230801_ArO.png  338 kB  | Show | Hide all | Show all
Attachment 3: bake2_230801_Closed_Open.png  540 kB  Uploaded Tue Aug 1 13:17:51 2023  | Hide | Hide all | Show all
bake2_230801_Closed_Open.png