177
Entry time:
Fri Jul 28 13:07:39 2023
Author:
Shane, Aiden
Type:
Update
Category:
Cleanroom
Subject:
cleanroom cleaning and particle count
[Aiden, Shane]
cleaning cleanroom and particle count
11:08 am: started particle count
zone 3:
0.3 u: 4198
0.5 u: 1080
1.0 u: 581
zone 4:
0.3 u: 1247
0.5 u: 623
1.0 u: 415
11:24 am: break for removal of vac chamber insulation (pictures attached), with counts before removal as seen above, and counts after removal/replacement for zone 3 as follows:
zone 3:
0.3 u: 2993
0.5 u: 415
1.0 u: 207
12:02 pm: began surface check and wipedown
12:13 pm: started vacuuming the floor
12:25 pm: finished vacuuming the floor
12:26 pm: started mopping the floor
12:32 pm: finished mopping the floor
12:32 pm: started cleaning the buckets
12:37 pm: started mopping with IPA wipes
12:43 pm: finished mopping with IPA wipes
12:45 pm: changed sticky floor mats
12:47 pm: started particle count
zone 3:
0.3 u: 2951
0.5 u: 290
1.0 u: 41
zone 4:
0.3 u: 290
0.5 u: 0
1.0 u: 0
Attachment 1:
image_50412289.JPG
1.612 MB
|
Hide
|
Hide all
|
Show all
Attachment 2:
image_50432769.JPG
1.955 MB
|
Hide
|
Hide all
|
Show all
Attachment 3:
image_67126017.JPG
1.525 MB
|
Show
|
Hide all
|
Show all
Attachment 4:
image_67163137.JPG
1.363 MB
|
Hide
|
Hide all
|
Show all
Attachment 5:
image_67168001.JPG
1.284 MB
|
Hide
|
Hide all
|
Show all
Attachment 6:
IMG_8427.jpeg
3.642 MB
|
Hide
|
Hide all
|
Show all
ELOG V3.1.3-7933898