ligo-ex ligo-ds
  Richardson Lab Experimental Log  Not logged in ELOG logo
Message ID: 126     Entry time: Fri Jun 16 17:40:28 2023
Author: Aiden 
Type: Update 
Category: Clean & Bake 
Subject: Clean and Bake Batch 14 
Cleaned and Baked the new and 2.75" copper gaskets for the vacuum system. Cleaned them with liquinox in the ultra sonic washer for 10 minutes. Then dried them with nitrogen. Then put the copper parts into the oven for the following steps; 1. Ramp to 100 degC in 15 minutes 2. Dwell at 100 degC for 30 minutes 3. Ramp to 175 degC in 30 minutes 4. Dwell at 175 degC for 24 hours 5. Turn off Also bagged and tagged the parts from Batch 13 and placed them into the clean room on the computer table.
Attachment 1: IMG_4441.jpg  76 kB  | Show | Hide all | Show all
Attachment 2: IMG_4440.jpg  121 kB  | Hide | Hide all | Show all
IMG_4440.jpg
ELOG V3.1.3-7933898